SPECIFICATIONS
•
111
Chapter 9
Zap
Amplitude
+1.3 V
DC
at pipette for chosen duration.
Duration
0.5-50 ms or Manual. Triggered by front-panel pushbutton. In Manual position ZAP
amplitude is maintained as long as pushbutton is depressed.
Command Potentials
Seal Test
5 mV command at line frequency (voltage clamp mode)
50 pA (current clamp mode,
β
= 1) or 500 pA (current clamp,
β
= 0.1).
External Commands
Two separate BNC inputs, one front switched, one rear switched.
Sensitivity
FRONT-SWITCHED:
20 mV/V in V-CLAMP, 2
÷
β
nA/V in I-CLAMP,
disabled in TRACK and I=0.
REAR-SWITCHED:
100 mV/V in V-CLAMP, 2
÷
β
nA/V in I-CLAMP,
disabled in TRACK and I=0.
Input impedance
10 k
Ω
. Inputs may be connected in parallel to increase sensitivity.