![Axon Axopatch 200B Скачать руководство пользователя страница 130](http://html.mh-extra.com/html/axon/axopatch-200b/axopatch-200b_theory-of-operation_3040439130.webp)
114
•
SPECIFICATIONS
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
Low Pass Bessel Filter
4-pole lowpass Bessel filter with five settings: 1, 2, 5, 10 and 100 kHz. Selected value sets
an analog voltage on FREQUENCY TELEGRAPH OUTPUT.
Leak Subtraction
Causes a signal proportional to the command to be subtracted from current record. Range:
100
β
M
Ω
to
∞
.
Telegraph Outputs
Gain
Takes
α
and
β
gain factors into account
I (mV/pA):
0.05
1
0.1
1
0.2
1
0.5
1
2
5
10
20
50
100
2
200
2
500
2
Vm (mV/mV):
N/A
N/A
N/A
0.5
1
2
5
10
20
50
100
200
500
Telegraph Output (V): 0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
Frequency
Filter Setting (kHz):
1
2
5
10 100
Telegraph Output (V):
2
4
6
8
10
Mode
I-CLAMP
I-CLAMP
Mode:
TRACK
V-CLAMP
I = 0
NORMAL
FAST
Scaled Output: I
I
V
m
V
m
V
m
Telegraph (volts): 4
6
3
2
1
1
Applicable for
β
= 0.1 only
2
Applicable for
β
= 1 only
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...