![Axon Axopatch 200B Скачать руководство пользователя страница 129](http://html.mh-extra.com/html/axon/axopatch-200b/axopatch-200b_theory-of-operation_3040439129.webp)
SPECIFICATIONS
•
113
Chapter 9
Blank Activate
Causes SCALED OUTPUT and I OUTPUT to hold their initial value for the duration of
the blanking pulse. Does not affect 10V
m
output.
Speed Test
Injects current into headstage input through a 1 pF capacitor. Injected current waveform is
the derivative of the voltage waveform applied at SPEED TEST input. For example, a
100 Hz 10 V
p-p
triangle wave will inject a 1 nA
p-p
square wave into the headstage input.
Signal Outputs
Scaled Output
Scaled and filtered by output control settings. Sample and hold pedestal compensation.
Output is I (
αβ
mV/pA) when in V-CLAMP or TRACK (I=0) mode. Output is V
m
(
α
mV/mV) when in I-CLAMP mode. BNCs on front and rear panels are identical.
I
Pipette current. Rear switched gain of either
β
mV/pA or 100
β
mV/pA; fixed filter: 10 kHz
3-pole Bessel. Output does not benefit from sample and hold pedestal compensation.
10V
m
Membrane potential at x10 gain. Junction potentials removed.
Output Controls
Output Gain(
α
αα
α
)
10 values from 0.5 - 500. Affects SCALED OUTPUT only. Selected value sets analog
voltage on GAIN TELEGRAPH OUTPUT for reading by computer.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...