IMPORTANT INFORMATION
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
Disclaimer
THIS EQUIPMENT IS NOT INTENDED
TO BE USED AND SHOULD NOT BE
USED IN HUMAN EXPERIMENTATION
OR APPLIED TO HUMANS IN ANY
WAY.
Warning
IF THIS EQUIPMENT IS USED IN A
MANNER NOT SPECIFIED BY THE
MANUFACTURER, THE PROTECTION
PROVIDED BY THE EQUIPMENT MAY
BE IMPAIRED.
Power-Supply Voltage Selection and
Fuse Changing
Supply Voltage
The Axopatch 200B can be directly
connected to all international supply
voltages. The input range is from 100 to
240 V~. No range switching is required.
Changing the Fuse
The Axopatch 200B uses a 250 V~, T2A,
5 x 20 mm fuse.
In the event of fuse failure, disconnect
the power cord.
Before changing the fuse investigate the
reason for its failure.
To change the fuse:
1.
Disconnect the power cord.
2. Use a screwdriver or a similar device
to rotate the fuse holder
counterclockwise.
3. Replace the fuse with another fuse
of the same rating.
4. Reconnect the power cord.
Basic Equipment Setup and Safety
1. Supply and Earthing
Connections: Use the included IEC
power cord to connect the
instrument to a GROUNDED
power receptacle.
2. Mounting: Table or rack.
3. Assembly: The headstage connects
to the instrument through the rear
panel, 15 pin D-sub connector
marked "headstage".
4. Use: Do not operate this equipment
with covers or panels removed.
5. Cleaning: Wipe the headstage
connector with a damp cloth to
clean salt spills. Avoid spilling
liquids on the headstage.
The Teflon input connector should
be kept very clean. Effective
cleaning can be done by spraying
with alcohol or swabbing carefully
with deionized water. If possible,
avoid the use of Freon since it is
thought to be detrimental to the
environment.
Safe Environmental Conditions
1. Indoor use.
2. Mains supply fluctuations: not to
exceed ±10% of the nominal
voltage.
3. Temperature: between 5 ºC and
40 ºC.
4. Altitude: up to 2000 m
5. This instrument is designed to be
used under laboratory conditions.
Operate in a clean, dry
environment only. Do not operate
in a wet or damp environment.
Static Precautions
The headstage can normally be safely
handled. However, if you are in a
laboratory where static is high (
i.e.
, you
hear and feel crackles when you touch
things), you should touch a grounded
metal object immediately before touching
the headstage.
You should
not
switch off power to the
Axopatch 200B when handling the
headstage input since this will upset the
thermal equilibrium.
WARNING
Shipping the Axopatch 200B
The Axopatch 200B is a solidly built
instrument designed to survive shipping
around the world. However, in order to
avoid damage during shipping, the
Axopatch 200B must be properly
packaged.
In general, the best way to package the
Axopatch 200B is in the original factory
carton. If this is no longer available, we
recommend that you carefully wrap the
Axopatch 200B in at least three inches
(75 mm) of foam or "bubble-pack"
sheeting. The wrapped Axopatch 200B
should then be placed in a sturdy
cardboard carton. Mark the outside of
the box with the word FRAGILE and an
arrow showing which way is up.
We do not recommend using loose foam
pellets to protect the
Axopatch 200B. If the carton is dropped
by the shipper, there is a good chance
that the Axopatch 200B will shift within
the loose pellet packaging and be
damaged.
If you need to ship your Axopatch 200B
to another location, or back to the
factory, and you do not have a means to
adequately package it, Axon Instruments
can ship the proper packaging material
to you for a small fee. This may seem
like an expense you would like to avoid,
but it is inexpensive compared to the cost
of repairing an instrument that has
sustained shipping damage.
It is your responsibility to package the
instrument properly before shipping. If it
is not, and it is damaged, the shipper will
not honor your claim for compensation.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...