USE OF THE PATCH CLAMP – A TUTORIAL
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
to lower the pipette until the fractional resistance increase is optimal to promote sealing to
the cell. For many cells, a two-fold resistance increase is optimal (the size of the square
pulse on the oscilloscope decreases in half) but you will have to determine the optimum for
your particular cell type.
2 ms
250 pA
IN BATH
PUSHED AGAINST CELL
MORE PRESSURE
AGAINST CELL
GIGOHM SEAL
Figure 8.
Change in resistance while forming a seal.
At this point, apply suction to the tubing connected to the pipette holder. Some
investigators prefer to use mouth suction while others find that a 10 ml syringe works well.
With the latter approach, it is best to start with the syringe plunger pulled back to a volume
of 5-7 ml so that approximately this volume exists in the system before suction is applied.
Pulling the syringe back by about 1 ml from this starting volume will reduce the pressure
by 0.2 to 0.14 atmospheres. Since many other schemes of applying suction are used by
investigators, we do not provide here an exhaustive description. As suction is applied,
there should be a sudden increase in the resistance. Once a gigohm seal is established, the
rectangular current pulse will disappear entirely and be replaced by capacitance transients
in synchrony with the rising and falling edges of the command pulse (Figure 8, lowest
trace).
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...