6
•
FUNCTIONAL CHECKOUT
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
Test WHOLE CELL and PATCH configurations:
Figure 1.
Connections for testing Axopatch 200B and PATCH-1U model cell.
6) Remove the foil shield temporarily. Connect the black lead (2 mm pin at one end) to
the PATCH-1U model cell ground (2 mm socket at central position) then connect this
lead to the ground input of the headstage (1 mm pin).
7) Connect the model cell BATH connection to the headstage input. Insert the Teflon
collet of the model cell into the collar of the headstage (the fit will be snug). Make
sure that the collet is inserted all the way so that the connector pin is fully inserted into
the headstage input jack.
8) Shield the headstage and model cell with the foil. Reconnect the foil shield to ground.
9) Place the CONFIG. switch in the WHOLE CELL (
β
= 1) position. Turn the panel
meter switch to the I position. Zero the panel meter using the PIPETTE OFFSET knob.
If the meter cannot be easily zeroed within ±10 pA or if it constantly drifts, then the
foil shield is probably not correctly connected to ground.
10) Turn the panel meter switch to V
HOLD
/I
HOLD
and set the HOLDING COMMAND switch
to "+" and the toggle to "x1". Turn the HOLDING COMMAND potentiometer until
the meter reads 10 mV. Turn the panel meter switch back to I.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...