PRINCIPLES OF OPERATION
•
91
Chapter 8
For best results, the cell membrane resistance should be many fold higher than the
pipette resistance. This is normally the case for cells at rest carrying small drug-
activated or synaptic currents. However, during voltage activation the cell
membrane resistance could fall a hundredfold or more to values similar to or less
than the series resistance. In these cases it is probable that:
1)
There will be a significant error due to the voltage drop across the pipette.
This error is not obvious to the user because the patch clamp controls the
combined voltage drop across the pipette and the cell.
2)
The setting of the SERIES RESISTANCE and WHOLE CELL CAP.
compensation controls will become erroneous because it is based on the time
constant to charge the membrane capacitance before the change in membrane
resistance occurred. Since this time constant depends on the parallel value of
membrane resistance and the pipette series resistance, this error could
become substantial. The effect will be a larger transient at voltage levels that
activate the fall of membrane resistance.
If the cell input resistance becomes comparable to, or less than, the pipette
resistance, the whole-cell patch technique will probably not work. Note that
typical patch pipette resistances are in the range of 2 to 10 M
Ω
. In this case, it
would be preferable to use a discontinuous (chopped) single-electrode voltage
clamp, such as the Axoclamp, that will give more accurate results.
Pipette Offset
The PIPETTE OFFSET control is a ten-turn potentiometer used to add up to ±250 mV to
the pipette command potential (V
p
). in order to compensate for the total offset of the
liquid-liquid and liquid-metal junction potentials in the electrode and bath, and the offset of
the probe input amplifier. It is used at the beginning of each experiment to zero the pipette
current (I) when the electrode first touches the solution, and may be used occasionally
thereafter to manually adjust for any offset drift. The PIPETTE OFFSET control should
not be adjusted after formation of a tight seal between pipette and cell.
A description of how to use the PIPETTE OFFSET is given in the
Pipette Offset
Adjustment
section of the
TUTORIAL
.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...