USE OF THE PATCH CLAMP – A TUTORIAL
•
33
Chapter 3
If the seal was made using the tracking circuit, this circuit can now be turned off by
switching to the V-CLAMP mode. Turn off the SEAL TEST switch and turn the meter
switch to the I
RMS
position. Check the noise level on the meter and decide if the quality of
the seal and pipette coating are satisfactory. With the Axopatch 200B integrating
headstage, proper pipette coating, low loss pipette glass, and proper grounding and
shielding, it is possible to obtain noise levels in the 0.095-0.12 pA rms range if a high
resistance seal is obtained. Forgetting to turn off the SEAL TEST signal will result in very
high noise levels because of the noise contribution of the capacitance transients.
Turn the SEAL TEST signal back on and, if you plan to do pulsed experiments, proceed to
cancel the capacitance transients. This is done using the fast and slow PIPETTE
CAPACITANCE COMPENSATION circuits. First, turn the SLOW MAG control fully
counterclockwise so that there will initially be no compensation for a slow component.
Adjust the FAST MAG and FAST
τ
controls until the capacitance transient is minimized.
(
Note:
with some practice you will be able to adjust MAG and
τ
simultaneously and
decrease the time required to do the adjustments). If a slow component remains, minimize
its size using the SLOW MAG and SLOW
τ
controls. Turn off the SEAL TEST switch and
you are ready to record channel currents.
In some experiments, you may not require complicated pulse protocols and the application
of a simple DC voltage to the Axopatch 200B command line is all that is required. For this,
you can use the HOLDING COMMAND circuit included in the Axopatch 200B. Turn the
meter switch to V
HOLD
/I
HOLD
and the meter will display the absolute value of the voltage set
by the HOLDING COMMAND controls. This is true even with the HOLDING
COMMAND switch in the OFF position. This feature allows you to preset the voltage
without delivering it to the preparation until you are ready. Note that the HOLDING
COMMAND toggle gives you the choice of two ranges, x1 and x5. The voltage actually
delivered to the preparation will be positive or negative depending on the position you
choose for the HOLDING COMMAND switch (+ or -).
At this point, you should be seeing single-channel currents if channels exist in the patch.
Adjust the OUTPUT GAIN rotary switch to a position appropriate for the size of the
currents.
If you plan to use an external function generator or the output of a computer driven D/A
converter to provide command potentials, the signal should be connected to the EXT.
COMMAND INPUT BNC (front switched or rear-switched) on the back of the
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...