116
•
SPECIFICATIONS
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
PATCH-1U Model Cell
Emulates three experimental conditions:
BATH: 10 M
Ω
electrode resistor to ground. 4 pF pipette capacitance.
CELL:
10 M
Ω
electrode resistor connected to a 500 M
Ω
// 33 pF cell. 4 pF pipette
capacitance.
PATCH: 10 M
Ω
electrode connected to a 10 G
Ω
patch. 5 pF pipette capacitance.
MCB-1U Model Bilayer
Emulates a bilayer membrane. 10 k
Ω
resistor in series with a 100 pF capacitor.
Headstage Dimensions
Case is 0.7" x 0.75" x 4.2" (17.8 x 19 x 106.7 mm). A removable mounting plate 2.5 x 2.0
x 0.25" is supplied. Cable length is 10 feet (3 m).
Cabinet Dimensions
3.5" (89 mm) high, 19" (483 mm) wide, 12.5" (317 mm) deep. Mounts in standard 19"
rack. Handles included. Net weight 11.5 lbs (5.1 kg).
Supply Requirements
Line Voltage:
100 to 240 V
~.
Universal voltage input.
Line Frequency:
50-60 Hz.
Power
30 W
Fuse
0.5 A slow. 5 x 20 mm.
Line Filter:
RFI filter is included.
Line cord:
Shielded line cord is provided.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...