INSTRUMENT OPERATION
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
MODE TELEGRAPH OUTPUT
Provides a series of voltages that can be read by the computer to determine the
setting on the MODE switch. The voltages and their associated switch settings are
as follows:
I-CLAMP
I-CLAMP
Mode:
TRACK
V-CLAMP
I = 0
NORMAL
FAST
Scaled Output:
I
I
V
m
V
m
V
m
Telegraph (volts):
4
6
3
2
1
SCALED OUTPUT
Provides the current (filtered and scaled) in V-CLAMP and TRACK modes or the
membrane voltage (filtered and scaled) in I-CLAMP modes. This BNC is
duplicated on the front and back panels. On the front panel, two LEDs specify
whether the output is current (I) or membrane voltage (V
m
). The current gain is
α
β
mV/pA. The voltage gain is
α
mV/mV.
10 V
m
OUTPUT
Provides the membrane voltage multiplied by ten (x10) for use in current-clamp
experiments.
In V-CLAMP mode, 10 V
m
is ten times the
command
potential. If the pipette
current is zero, or if the series resistance correction is 100%, the command
potential is identical to the
actual
membrane potential. The actual membrane
potential is given by the following equation:
10 V
m
(100 -
CORR
)
Actual Membrane Potential (mV) =_______
_
________
I R
s
10
100
where
I
= pipette current in nA
CORR
= % correction set on SERIES RESISTANCE
CORRECTION control
R
s
= resistance in M
Ω
set on SERIES RESISTANCE control
In TRACK mode, 10 V
m
is ten times the
actual
potential.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...