INSTRUMENT OPERATION
•
51
Chapter 6
Seal Test
The SEAL TEST command generator provides a convenient source of test pulses to be
used during seal formation. The step is gated by the internal line frequency oscillator; its
duty cycle is between 40% and 60%. In voltage clamp mode, the step size is 5 mV; in
current clamp, it is 50 pA (
β
= 1) or 500 pA (
β
= 0.1).
Current Clamp
The Axopatch 200B can be used in current-clamp mode similarly to a conventional
microelectrode amplifier. Series Resistance compensation is functional in this mode.
There are three current-clamp modes. I=0 for voltage recording without an applied current
command, I-CLAMP NORMAL for use with pipette resistances greater than 1 M
Ω
and
I-CLAMP FAST for use with pipette resistances greater than 10 M
Ω
. The accuracy of the
response can be improved by using the FAST and SLOW ELECTRODE CAPACITANCE
COMPENSATION as well as the SERIES RESISTANCE COMPENSATION controls.
Membrane Potential
In voltage-clamp mode, V
m
is simply equal to the command potential. In current-clamp
mode, V
m
is not clamped and automatically becomes the voltage needed to allow the
commanded current to flow.
Note:
To avoid introducing errors, you
should not
change the PIPETTE OFFSET control
after you switch from V-CLAMP to I-CLAMP.
Whole-Cell Current Clamp
Current-clamp can be used from the very beginning of the experiment, or you can switch
into current-clamp mode at any time (see
Current Clamp
section of the
TUTORIAL
).
However, if whole-cell recording has been established, special care must be taken when
switching between voltage-clamp and current-clamp modes because in voltage-clamp mode
the EXT. COMMAND sensitivity is 20 mV/V and the HOLDING COMMAND sensitivity
is 20 mV/turn (for x1; 100 mV/turn with the toggle set to x5). In current-clamp the EXT.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...