Routine
Maintenance
3
STERRAD
®
Sterilization System Routine Maintenance Guide
19
The Y changes to N automatically when a PM is due. The level of PM
due is indicated on the printout after a cycle has been completed, and on
the display in the idle state. You can reset the N to Y by pressing OPEN
DOOR; XXXX automatically changes to the level 1 or 2 PM service
interval in cycles. If no PM is due, then the number of cycles until the
next level 1 or 2 PM is displayed.
1. Press
OPEN DOOR
to store the data, the total machine cycles, and
print the information while exiting the Edit mode. The maximum
storage capacity is 10 PM service records. When the 11th PM service
record is entered, the system deletes the oldest record, re-sequences
the existing records and enters the new record in the 10th position.
The system prints the PM service history when it is updated.
2. Press
CANCEL
to transfer to the PM Level 1 Cycles Interval menu.
PM Level 1 Cycles Interval
The following is displayed:
PM 1 CYCLES = XXXX
OPEN(chng)-CLOSE(accept)
XXXX is the current Level 1 PM cycles service interval.
1. Press
CLOSE DOOR
to move the cursor to each digit in turn,
starting at the rightmost digit, moving left, and wrapping back to the
rightmost digit.
2. Press
OPEN DOOR
to increment the digit by 1 ranging from 0 to 9.
The Level 1 PM cycles can be programmed to a value between 500
and 2000 cycles, with a default value of 1500 cycles.
3. Press
CANCEL
to transfer to the PM Level 2 Cycles Interval menu.
PM Level 2 Cycles Interval
The following is displayed:
PM 2 CYCLES = XXXX
OPEN(chng)-CLOSE(accept)
XXXX is the default PM level 2 cycles service interval. The default PM
level 2 cycles interval is 2 times that of the PM level 1 cycles interval.