3
Routine Maintenance
26
STERRAD
®
Sterilization System Routine Maintenance Guide
As Needed Maintenance
The following items are serviced on an “as needed” basis and should be
checked at both the Level 1 and Level 2 PM intervals.
•
Change the cassette collection box.
•
Replace the injector valve vaporizer plate.
•
Replace the printer ribbon rolls or cartridge.
•
Replace the printer paper.
•
Reset the time and date display.
•
Set the pressure units.
•
Use the history menus.
•
Clean the sterilizer.
•
Master Reset
Changing the Cassette Collection Box
The cassette collection box can hold 30 used cassettes. The cassette box
must be changed after 145 cycles. The customer is instructed to change
the vaporizer plate every time a new cassette collection box is needed.
If the cassette collection box is not replaced at the proper cycle intervals,
the sterilizer does not run any more cycles and the paper printout shows:
CHANGE COLLECTION BOX
NO FURTHER CYCLES CAN BE RUN.
If this occurs, change the cassette collection box immediately. At the
same time, change the vaporizer plate.