Routine
Maintenance
3
STERRAD
®
Sterilization System Routine Maintenance Guide
23
1. Reinstall the panels and reconnect the cassette door assembly cable.
2. Return the door to the normal position. Clean the rear surface of the
door with deionized water.
3. Close the chamber door. Close the storage drawer.
Clean Injector Pump Platen Surface
1. Clean the surface of the injector pump platen with deionized water
and a lint free cloth.
PM Main Menu
PM WARNING = ON (or) OFF OPEN(chng)-
CANCEL(next)
ON or OFF indicates the state of the Planned Maintenance warning
feature. When this feature has been turned OFF, no messages related to
performing planned maintenance are generated. Press OPEN DOOR to
toggle the warning between ON and OFF. Press CANCEL to accept the
choice displayed.
If the warning feature is ON, pressing CANCEL gives different options
for selecting the PM interval. If this feature is OFF, Press
CANCEL
to
transfer to the Serial Number Edit Menu.
The PM warning is set to ON by default, whenever a loss of battery
power occurs or the number of total machine cycles is reset to zero.
PM Level 1 (2) Edit Menu
The following is displayed:
PM Y (or) N; NEXT PM = XXXX L(X)
OPEN(reset)-CANCEL(next)
“Y” is yes; PM was done. “N” is no; PM was not done. XXXX is the
number of cycles before the PM is due, or the month in which the PM is
due and (X) is level 1 or level 2. If PM is not due on the basis of either
months or cycles, XXXX is in cycles. If both months and cycles are due,
or both are past due, XXXX is also in cycles. XXXX is displayed in
months, if months are due, and cycles are not due. Months are also
displayed if months are past due, while cycles are less than past due. If
the number shown is a negative number, then it is the number of cycles
past due.