STERRAD
®
Sterilization System Routine Maintenance Guide
11
Chapter 3.
Routine Maintenance
This chapter presents the routine maintenance, or PM (planned
maintenance) procedures that are performed on the STERRAD
®
100S
Sterilizer. There are two maintenance intervals: PM Level 1 and PM
Level 2. These intervals are tracked by the system and the user is
notified when the PM is due. The PM Level 1 is performed every 1500
cycles or 6 months whichever comes first. PM Level 2 is performed
every 3000 cycles or 1 year, whichever comes first.
In addition, there are a number of “as needed” maintenance activities to
be performed. These are presented near the end of this chapter.
WARNING! ONLY EXPERIENCED, ASP-TRAINED
TECHNICIANS SHOULD REPAIR OR
ADJUST THIS UNIT. REPAIRS AND
ADJUSTMENTS SHOULD ONLY BE
ATTEMPTED BY EXPERIENCED
TECHNICIANS WHO ARE FULLY TRAINED
TO MAINTAIN AND REPAIR THE STERRAD
STERILIZER.
USE OF UNAUTHORIZED PARTS FOR
MAINTENANCE OR REPAIR COULD CAUSE
PERSONAL INJURY, RESULT IN COSTLY
DAMAGE OR UNIT MALFUNCTION, AND
VOIDS THE WARRANTY.
Caution: Do not leave the sterilizer unplugged or
turned off for longer than 24 hours. If the
sterilizer must be turned off for longer than
24 hours, call the ASP Customer Care
Center for instructions.