User Manual
KDC Series II DC Power Supplies
83
7.3 Routine Calibration
Setup:
Connect the test equipment to the power supply as shown in Figure 7-1. The DMM can be
connected to the common output terminal at the rear of the unit.
Never connect the load wire to the SENSE terminal at the Rear Panel terminal strip. The load
must be connected to the POWER bus bars and a jumper between the POWER and SENSE
terminal. Always connect the external DVM to the SENSE terminal.
L
R
SHUNT
CURRENT
VOLTAGE
DC SUPPLY
(rear view)
POS.
NEG.
Measure output
voltage at sense
connection.
Figure 7-1: Test Equipment Hookup for Routine Output and Measurement Calibration