User Manual
88
KDC Series II DC Power Supplies
7.5.2
Current Measurement Calibration
Current measurement calibration applies to both average current and peak current measurement
functions.
Offset
1. Set up the DC supply as described under output current offset calibration.
2. Read the measured output current using an external shunt and DMM.
3. Enter the value measured in the MEAS CAL, MCURR OFST screen using one of the
shuttles.
4. Press the ENTER key.
5. The DC supply measurement systems will take a few seconds to set its measurement
calibration coefficient based on the reference measurement value provided.
Full Scale
1. Set up the DC supply as described under output current full-scale calibration.
2. Read the measured output current using an external shunt and DMM.
3. Enter the value measured in the MEAS CAL, MCURR F/S screen using one of the shuttles.
4. Press the ENTER key.
5. The DC supply measurement systems will take a few seconds to set its measurement
calibration coefficient based on the reference measurement value provided.
7.5.3
Power Measurements
All other measurements except voltage and current made by the DC supply are mathematically
derived from the voltage and current measurements. As such, no other measurement
calibrations are required.
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...