User Manual
86
KDC Series II DC Power Supplies
Full Scale - External 5 V Input
1. Change the DC supply control mode for CURR to REF EXT5V from the CONTROL screen.
2. Use a small precision DC supply or DC calibrator to apply a 5 V DC input to the RPI input.
3. Monitor the output current using an external current shunt and DMM
4. Select the OUTP CAL, RPI 5V screen.
5. Increment or decrement the RPI 5V calibration coefficient while monitoring the output current
until the measured current matches the programmed setting.
Full Scale - External R Input
1. Change the DC supply control mode for CURR to REF EXTRES from the CONTROL screen.
2. Use a suitable resistor between the Remote R current programming input.
3. Monitor the output current using an external current shunt and DMM
4. Select the OUTP CAL, RPI RES screen.
5. Increment or decrement the RPI RES calibration coefficient while monitoring the output
current until the measured current matches the programmed setting.