User Manual
KDC Series II DC Power Supplies
37
ENTRY
DESCRIPTION
BUS
Bus mode. The transient system is started by a
bus command or a group execute trigger (GET).
EXT
External mode. The transient system is started
by a user-provided external trigger signal.
TRAN STEP
Indicates the transient system execution mode. Available
modes are:
AUTO
When triggered, the transient system will
automatically execute each list point sequentially
without waiting for a trigger between list points.
This execution is paced by the dwell time set for
each data point.
ONCE
When triggered, the transient system will
execute the first list point and wait for a new
trigger once the dwell time expires. This allows
triggered execution of each step in the transient
list.
List parameters:
VOLT
Step #
Voltage set point. See Table 4-2.
VSLEW
Step #
Voltage slew rate in V/sec
Minimum setting: 0.001 V/s
Maximum setting: 1E6 V/s or Vrange * 1000
whichever is smaller.
Resolution: 1 msec
CURR
Step #
Current set point. See Table 4-2.
CSLEW
Step #
Current slew rate in A/sec
Minimum setting: 0.001 A/s
Maximum setting: 1E6 A/s or Irange * 1000
whichever is smaller.
Resolution: 1 msec
DWELL
Step #
Dwell time in seconds.
Range is 0.001 to 9E4 sec
Resolution: 1 msec
TTLT
Step #
ON:
Generates an output trigger at this step.
OFF:
No output trigger.
Transient List points data entry method.
Transient list points are numbered sequentially from 0 through 99 and executed in this order.
Each list point or list entry has 6 parameters as shown in the table above. To enter list point data,
the right hand knob (CURRENT) is used to increment or decrement the parameter value. The left-
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...