User Manual
44
KDC Series II DC Power Supplies
Output Calibration Menu
The output calibration menu can be used to perform routine calibration of the dc supply output.
The recommended calibration interval is 12 months. The output calibration menu allows
calibration of both the output and the external analog inputs if installed.
Note: Refer to chapter 7 for details on routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
ENTRY
DESCRIPTION
VIEW ANGLE
-10 to +10
LCD viewing angle adjustment.
CAL PWORD
V range
Calibration password required to access all
calibration screens. The calibration password
is the voltage range value. Thus, for a KDC80-
187, the cal password is 80. The password
can be set using the voltage or current knob.
Analog I/O option Calibration Menus
RPV 5V
± 9999
Calibration coefficient for full-scale voltage 5V
RPV analog input
RPI 5V
± 9999
Calibration coefficient for full-scale current 5V
RPV analog input
RPV 10V
1
± 9999
Calibration coefficient for full-scale voltage
10V RPV analog input.
RPI 10V
2
± 9999
Calibration coefficient for full-scale current 10V
RPV analog input.
RPV RES
± 9999
Calibration coefficient for full-scale voltage
resistive analog input
RPI RES
± 9999
Calibration coefficient for full-scale current
resistive analog input
1
When using 10V input, it is recommended to ground the 5V input to avoid noise pickup.
2
When using 10V input, it is recommended to ground the 5V input to avoid noise pickup.
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...