User Manual
46
KDC Series II DC Power Supplies
4.2.9
Measurement Calibration Menu
The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months.
Note: Refer to chapter 7 for details on routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
ENTRY
DESCRIPTION
VIEW ANGLE
-10 to +10
LCD viewing angle adjustment.
CAL PWORD
V range
Calibration password required to access all
calibration screens. The calibration password
is the voltage range value. Thus, for a KDC80-
187, the cal password is 80. The password
can be set using the voltage or current knob.
Measurement Calibration Screens
MVOLT F/S
± 9999
Calibration coefficient for full-scale voltage
measurement.
MCURR F/S
± 9999
Calibration coefficient for full-scale current
measurement.
MVOLT OFST
-128 - + 127
Calibration coefficient for voltage
measurement offset.
MCURR OFST
-128 - + 127
Calibration coefficient for current
measurement offset.
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...