Following calibration screens are used to calibrate the voltage and current measurements.
7.5.1
Voltage Measurement Calibration
Offset
1. Set up the DC supply as described under output voltage offset calibration.
2. Read the measured output voltage from the DMM.
3. Enter the value measured in the MEAS CAL, MVOLT OFST screen using one of the shuttles.
4. Press the ENTER key.
5. The DC supply measurement systems will take a few seconds to set its measurement
calibration coefficient based on the reference measurement value provided.
Full Scale
1. Set up the DC supply as described under output voltage full-scale calibration.
2. Read the measured output voltage from the DMM.
3. Enter the value measured in the MEAS CAL, MVOLT F/S screen using one of the shuttles.
4. Press the ENTER key.
5. The DC supply measurement systems will take a few seconds to set its measurement
calibration coefficient based on the reference measurement value provided.