User Manual
50
KDC Series II DC Power Supplies
4.3.3
Control the Output
The Output can be disabled or enabled using the ON/OFF button as follows:
1. Pressing the OUTPUT button when the output LED is on will turn off the DC supply output.
The programmed voltage setting will remain at the last program value.
2. Pressing the OUTPUT button again will engage the output and the output will revert to the last
programmed value.
4.3.4
Measure Peak Inrush Current
To measure the inrush current of an EUT, proceed as follows:
1. Turn of the supply output using the OUTPUT ON/OFF button and connect the EUT.
2. Press the MENU key and select the PROGRAM entry. Press ENTER to bring up the
PROGRAM screen.
3. Set the output to zero volt and enable the output using the OUTPUT ON/OFF button.
4. Use the SET key as described in section 4.3.1 above to change the output voltage to the
nominal supply voltage of the EUT.
5. Press the MEAS key to bring up the measurement screen, including the peak current
measurement.
6. Read the inrush current from the AP field in the lower right corner of the screen.
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...