User Manual
KDC Series II DC Power Supplies
41
4.2.8
Configuration Menu
The configuration menu may be used to configure various aspects of the dc supply such as the
serial port, GPIB bus (if installed) and the power on settings of the supply.
ENTRY
DESCRIPTION
ADDRESS
0 - 31
Sets the selected IEEE / GPIB bus address for
the optional IEEE/GPIB interface.
Factory default is address 1. The left knob
(VOLTAGE) can be used to scroll through the
0 through 31 address range. Do not use
address 0 as this address is typically reserved
for the GPIB controller.
BAUD RATE
9600
19200
38400
57600
115200
Sets the baud rate for the RS232C serial
communications port.
Factory default is 38400 baud. Available
settings are 9600 through 115200 baud. The
right knob (CURRENT) can be used to scroll
through these selections.
POWER ON
REG 0
– 7,
DEF
Selects the non-volatile register to be recalled
automatically at power-on.
Factory default is DEF, which recalls the
factory settings. Available parameters are
REG 0 through REG 7, which recalls any of
the user-defined setups, or DEF for factory
default.
Note that to use a register for power-on
default, the contents of the register must be
programmed first. See section 4.2.7. If an
empty register is selected, the DC supply will
revert back to DEF (factory setting).
Note: When operating the KDC power supply
as an auxiliary unit in a multi box configuration,
do not recall settings that were saved while the
KDC was in Master (normal) mode as this may
invalidate the auxiliary settings. This is
especially important if a register is assigned to
the power on default setting.
Содержание KDC 100-100
Страница 2: ......
Страница 10: ......
Страница 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Страница 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Страница 82: ......
Страница 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Страница 84: ...User Manual 74 KDC Series II DC Power Supplies Figure 6 2 DC Power Supply Block Diagram with IF option ...
Страница 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...