CHAPTER 8
Specifications ROXY Exceed
85
C H A P T E R 8
Specifications ROXY Exceed
Environmental, dimensions, weight & power requirements
Working temperature
Storage temperature
10 - 35°C (indoor use only)
–25 - +50°C
Humidity
20 - 80% RH
Safety and EMC
CE, UL, CSA, CB Scheme
Equipment class
1
Installation category
II
Pollution degree
2
Dimensions
43 (D) x 22 (W) x 44 (H) cm = 16.9” (D) x 8.7”
(W) x 17.3” (H)
Weight
14.4 kg (32 lbs) without flow cell and column
(SSC version)
Installation
Install upright on flat & smooth surface, keep
space under the potentiostat free.
Power requirements
100-240 VAC, 50/60 Hz, 260 VA, auto-sens-
ing
Mains fuse
2.5 AT / 250V, 5x20 mm, IEC 60127-2
For safety reasons, any other internal fuse or
circuit breaker is not operator accessible, and
should be replaced only by Antec authorized
personnel. Only use manufacturer-supplied
fuses.
For optimum analytical performance it is recommended that the
ambient temperature of the laboratory be between 20-25 °C and be held
constant to within ± 2 °C throughout the entire working day. Note: that
for optimal performance of the oven the oven temperature should be
set at least 7 degrees higher as ambient temperature.
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...