Chapter 7
Optimization of working potential
75
Fig. 31. Example of the mass spectra of 10
mol/L amodiaquine oxidized
at a glassy carbon working electrode, at a scan speed of 10 mV/s.
Optimization using a voltammogram
Sometimes, when interfering peaks appear in the chromatogram, it is pos-
sible to optimize the method with regard to selectivity. If the interfering
compound has a higher oxidation potential, a working potential is chosen
that gives the best selectivity, i.e. the largest difference in peak height. In
the example of Fig. 32 the selectivity for compound X is improved consid-
erably by decreasing the potential to E
2
or E
1
. Obviously, if compound Y is
the compound of interest, optimization of selectivity in this way is not pos-
sible and the chromatography has to be optimized.
Electrochemical detection differs from most other LC detection methods in
that a reaction takes place in the detection cell. Due to reaction kinetics an
increased temperature speeds up the oxidation/reduction reaction. How-
ever, this not only holds for the analyte but also for the background current
and possible interferences. An elevated temperature will therefore not au-
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...