1. Before switching ON the cell, make sure that the buffer contains
sufficient electrolyte (buffer ions). A stable working conditions will
never be obtained if the cell is switched ON with only water or
another non-conducting mobile phase. Also be sure that no air
bubbles are trapped in the electrochemical cell.
2. Set the cell potential (see chapter 9 for optimization of the
potential), switch ON the flow cell to oxidize or reduce the sample.
Your system is now ready for use.