Chapter 5
Measurement modes
67
Fig. 24. Potential steps in pulse mode. During t1 the measurement poten-
tial is applied, the actual current measurement occurs during ts. Steps t2,
t3 and t4 are for regenerating the electrode. This process repeats itself
continuously as soon as the cell is on.
In pulse mode up to five potential steps can be defined. The sampling in-
terval (period at which the current is measured) is applied at the end of E1.
Pulse mode - 2
In the ROXY Exceed an extended pulse mode, pulse mode - 2, is available
when operating the instrument via a PC using Dialogue Elite or Chro-
meleon CDS software. The pulse mode - 2 is not available in manual oper-
ation using keyboard/display. In the extended pulse mode it is possible to
program a multi-step waveform, with up to 30 time, potential (t,E) coordi-
nates, see figure below. Furthermore, the measurement time interval in
which the current is measured,
marked by a ‘Begin’ and ‘End’ marker, is
free programmable in the pulse table.
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...