Grounding kit
For the ROXY potentiostat (pn) an optional grounding kit pn 250.0035 is
available.
W
ith a ROXY EC system (pn 210.0070) or ROXY EC/LC system (pn
210.0080C) this part is provided with the systems. An ESI interface of
an MS is usually operating at high voltages of typically 3
– 5 kV. In
cases where the inlet of the ESI-MS is not grounded, the grounding
kit (pn 250.0035) must be used. If not used it may lead to damage of
the ROXY Exceed potentiostat.
For detailed installation information please consult the relevant installation
documentation of the ROXY EC or EC/LC system or contact Antec
support.
Connect to inlet
of ESI-MS
Connect to outlet tubing
of the reactor cell
Connect to grounding
stud on the rear panel of
the ROXY potentiostat
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...