Chapter 7
Optimization of working potential
73
As can be seen in both Fig. 27 and Fig
. 29
, when the working potential is
increased the electrochemical reaction is enhanced hence the signal in-
creases. At a certain potential the I/E curve flattens. All analyte molecules
that reach the working electrode are converted at such a high rate that the
analyte supply becomes the limiting factor. At the working electrode sur-
face a stagnant double layer exists, where molecular transport takes place
by diffusion only. Therefore, the current at (and beyond) this potential is
called the
diffusion limited current
.
Fig. 29. Scanning voltammetry of 1.0 µmol/L norepinephrine (A) at a
glassy carbon working electrode, at a scan speed of 10 mV/s. Scan (B) is
the blank solvent.
In practice the choice of the working potential is a compromise between
sensitivity, selectivity and reproducibility or the yield in desired metabolite
formation (in ROXY Ec applications). In the example of Fig. 27 a working
potential (E
1
) of 0.8 V is chosen.
0.0
0.4
0.8
1.2
0
5
10
B
Ic
ell
(nA)
Ecell (V)
A
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...