CHAPTER 9
Rear panel I/O
101
Fig. 51. Top: CONFIG screen. Bottom: ROXY Exceed signal processing
from electrochemical flow cell to output. R is a selectable I/E resistor of
100M
Ω, 10MΩ, 100KΩ, 1KΩ, 100 Ω and 10 Ω in the I/E converter circuit.
The output parameter Vout source can also be selected/set using Dialogue
Exceed. Go to the device settings menu and Acquisition tab and set the
analogue output to DAC or I/E converter. The instrument is by default set
to DAC. For regular measurements it is advised to use the DAC output.
Fig. 52. Dialogue Elite Acquisition tab analog Output setting.
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...