
CHAPTER 10
Troubleshooting
115
No response, no product in MS
Possible cause
Remedy
No power
Check line voltage setting, plug in power
cord
Power switch off
Turn this switch ON (at the rear panel)
Faulty fuse
Replace fuse
Cell disconnected, or
switched off
Check connection
WE contact problem
uPrepcell: check the spring construction in
the WE contact, should be flexible. If not: re-
place.
Air or gas bubbles in cell
Read filling instructions, check for bubbles at
the outlet
Conditioning of WE
Follow conditioning instructions
Fouled WE
Clean WE
Saturation of output (overload)
Possible cause
Remedy
Damaged REF
Check with spare REF, replace if necessary
Damaged WE
Replace cell block
Cell incorrectly connected
Check connections (REF: black, WE: red,
AUX: blue)
Normal behavior
With high substrate concentrations it can be
normal behavior. Check with solvent (without
substrate) if current is at expected value and
in scale. Increase range if possible.
Erratic cell cell current value or offset after current compensation
In case an Autozero, Offset or a change of range is initiated on an instable
or changing current signal (baseline), this may lead to the display of an er-
ratic current value or erratic offset in current. This is expected behavior. An
Autozero, offset or change of measurement range will trigger an internal
calculation procedure in the detector based on the actual cell current. This
is an iterative process and the cell current is measured several times dur-
ing the calculation process. In case the cell current is not stable during the
calculation process this might lead to errors.
Do not execute an autozero, offset or range change when the cell
current is not stabile or the baseline did not stabilize.
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...