CHAPTER 9
Rear panel I/O
103
Table VI. ROXY Exceed I/E converter resistors
Current Ranges
I/E Resistor (
Ω)
1nA, 2nA, 5nA, 10nA
100M
20nA, 50nA, 100nA
10M
200nA,500nA, 1µA, 2µA, 5µA, 10µA
100K
20µA, 50µA, 100µA, 200µA, 500µA, 1mA
1K
2 mA, 5 mA, 10 mA
100
20mA, 50mA, 100mA
10
In the potentiostat accessory kit a dedicated Output cable, D9 male
–
open, 2m (pn 250.0128A) is supplied. It is advised to use this manufac-
turer-supplied cable for this type of measurements.
For reference, the layout of the analog data out connector is shown in the
table below.
Table VII. Analog data output connector lay-out.
Pin
Layout
1
V
out 1
2
V
out 2
3
V
out 3
4
V
out 4
5
V
out 5
6 - 9
GND
To measure for example the analog signal of cell 1 with an external A/D
converter:
•
Connect the signal wire, lead of pin 1 (V
out 1
), to the analog
measurement channel of the A/D converter.
•
Connect the GND wire, lead of pin 6 (or 7-9), to the corresponding
analog ground connection of the A/D converter.
Digital I/O connector
The potentiostat has one 25-pins digital I/O connector which enables con-
trol of (or by) external equipment. The IO connector contains 18 TTL con-
tacts (5 outputs and 13 inputs, 3.3V CMOS logic), 2 RELAYS (contact clo-
sure) and 1 ground (GND connection).
Содержание ROXY Exceed
Страница 6: ...6 211_0010 ROXY Exceed user manual rev 04...
Страница 14: ...14 211_0010 ROXY Exceed user manual rev 04...
Страница 38: ...38 211_0010 ROXY Exceed user manual rev 04...
Страница 40: ...40 211_0010 ROXY Exceed user manual rev 04 Overview of ROXY Exceed screens DC mode...
Страница 41: ...Chapter 4 ROXY Exceed control 41 Pulse mode...
Страница 42: ...42 211_0010 ROXY Exceed user manual rev 04 SCAN mode...
Страница 43: ...Chapter 4 ROXY Exceed control 43 CONFIG menu...
Страница 44: ...44 211_0010 ROXY Exceed user manual rev 04 DIAG menu...
Страница 92: ...92 211_0010 ROXY Exceed user manual rev 04...
Страница 112: ...112 211_0010 ROXY Exceed user manual rev 04...
Страница 120: ...120 211_0010 ROXY Exceed user manual rev 04...
Страница 126: ...126 211_0010 ROXY Exceed user manual rev 04 Working environment safety 10...