Troubleshooting
SPECTRO ARCOS FHX3X — 08.02.2022
— 65
8 Troubleshooting
Notice! Damage to instrument!
Improperly performed repairs may damage the instrument.
Do not repair the instrument yourself. Always notify your immediate
supervisor or contact your Service Center.
Malfunction
Remedy
Argon pressure too low
Check the argon supply line for leaks and pressure as verify
the remaining amount of argon.
Auxiliary gas flow
deviation
The set flow rate cannot be reached.
Check the argon pressure and the plasma torch for leaks
or blockages and whether all connections have been
made correctly.
Plasma gas flow
deviation
The set flow rate cannot be reached.
Check the argon pressure and the plasma torch for leaks
or blockages and whether all connections have been
made correctly.
Plasma stopped (Dmb
timeout)
The generator could not establish communication with the gas
control unit.
Turn off the instrument and restart it again. If the problem
persists, please contact the SPECTRO Service
Department.
Housing extraction
system error
Extraction rate of the instrument (generator) is below the
minimum rate.
Check the extraction rate and adjust it to the values
required in the operating manual.
Communication with the
gas control system
cannot be established
The generator could not establish communication with the gas
control unit.
Turn off the instrument and restart it again. If the problem
persists, please contact the SPECTRO Service
Department.
Gas control system
deactivated
The generator could not establish communication with the gas
control unit.
Turn off the instrument and restart it again. If the problem
persists, please contact the SPECTRO Service
Department.
The instrument’s door is
open
Close the door in order to start the plasma.