Operation
— 36
SPECTRO ARCOS FHX3X — 08.02.2022
The solution must also contain:
20 mL/L HCl
20 mL/L HNO
3
The deviation of the specified element concentrations must not exceed 0.5%.
Greater deviations may adversely affect the analytical performance.
II. Organic
applications:
For organic applications, a one-time spectrum calibration with the aqueous ICAL
solution must first be performed. Subsequently, the relevant reference spectra
must be recorded. For a detailed description, refer to the online help of the
instrument.
Commercial solutions for the spectrum calibration, optimization or equipment
check are available from your local SPECTRO organization.
In order to proceed with the spectrum calibration, proceed as follows:
1. Start the plasma.
2. Flush the sample introduction system for approx. 10 minutes.
Use distilled water for aqueous applications.
3. Select the spectrum calibration symbol in the navigation tool bar.
At the bottom of the window, the current calibration and the status of the
calibration are displayed. Click the “Start Spectrum Calibration” button in
order to calibrate the optics.
The information box communicates whether the calibration has been
completed successfully or if errors occurred.
The optics is now calibrated.