Maintenance
SPECTRO ARCOS FHX3X — 08.02.2022
— 45
7.3
Cleaning the load coil
Warning! Hot surface!
The plasma torch, plasma interface, and load coil may be very hot.
Let the components cool for 10 to 15 minutes before commencing work on
them.
When working on the plasma torch, plasma interface, and induction coil,
wearing protective gloves and safety goggles is mandatory.
Warning! Risk of injury!
Glass parts may break when handled incorrectly and cause cutting injuries.
When working on the sample introduction system, wearing protective gloves
and safety goggles is mandatory.
If glass is broken, a suitable container (bucket) must be used to collect the
larger pieces first. In order to remove small parts and glass fragments, a
suitable vacuum must be used.
The load coil may corrode because of the sample vapors.
In order to clean the coil, proceed as follows:
1. Switch off the instrument.
2. Remove the Plasma torch; see page 41.
3. Use a small brush (e.g. a toothbrush) and a grinding paste (e.g. toothpaste) to
clean the load coil.
Do not bend the load coil.
4. Remove the grinding paste completely from the load coil.
5. Reinstall the components.
The load coil is cleaned.