Technical data
— 24
SPECTRO ARCOS FHX3X — 08.02.2022
Description
Value
Argon gas connection
Inlet pressure during operation
7 bar (102 psi)
It is imperative not to exceed a maximum Ar
inlet pressure of 8.0 bar (116 psi).
Quality
≥
4.6 (99.996 %)
Consumption
≤
25 L/min
Oxygen supply for auxiliary gas (optional; for organic applications only)
Inlet pressure during operation
4.0 bar (58 psi)
Quality
≥
4.5 (99.995 %)
Flow rate
The flow rate depends on the application.
Extraction (continuous)
Plasma torch box
Dependent on ambient temperature
18 to 25 °C = 200 to 250 m³/h; 118 to 150 CFM
> 25 °C = 250 to 300 m³/h; 150 to 177 CFM
Generator extraction or without
extraction
250 – 300 m³/h; 150 – 177 CFM or blowing into
the room without obstruction