7.9.2
Cleaning the sampler
In order to clean the sampler, proceed as follows:
1. Unscrew the sampler (2) from
the OPI (1), using the OPI
tool (3).
Hold the front side of the
OPI facing down. This
prevents dirt particles from
falling into the OPI.
Fig. 27
2. Clean the sampler under running water using a nailbrush.
When heavy contamination must be removed, a suitable acid may be
used. Then flush the sampler thoroughly with distilled water.
3. Clean the sampler opening, e.g., using a pipe cleaner.
4. Let the sampler dry using a jet of air or argon.
5. Reinstall the sampler (2).
Cleaning of the sampler is now complete.