Maintenance
SPECTRO ARCOS FHX3X — 08.02.2022
— 43
7. Clean the components of the
plasma torch.
Place the torch (3) in aqua
regia (a mixture of one part
concentrated nitric acid and
three parts concentrated
hydrochloric acid).
Hazardous substance!
Observe the
corresponding
information on
hazardous materials.
Let the acid/cleaning
solution act for approx. 15
minutes.
Subsequently, rinse the
torch with distilled water and
let it dry completely.
Fig. 15
8. Use a lint-free cloth in order to clean the adapter; subsequently, proceed with
a visual inspection.
9. Re-mount the plasma torch.
The same steps—carried out in reverse order—are used for mounting
the plasma torch, see point 5
Push the torch (1) into the
adapter (2) until the plasma
gas connection is concentric
with the recess in the
adapter (3).
There is a clearance of
±1 mm to optimize the
torch’s height.
Reconnect the hoses (A =
Axillary, C = Plasma gas).
Fig. 16
Содержание 76004567