viii | Model 5100 HD Analyzer
Secondary Process Seals Requirement
A”This instrument is supplied with a
primary process seal only
(single seal
device) in accordance with ANSI/ISA 12.27.01-2003. Proper installation of
this instrument requires a
secondary process seal
if failure of the primary
process seal could allow process fluids to enter the conduit or field wiring
system. The maximum process gas pressure and temperature that will be ap-
plied are 10 psig (170KPa) and 140F (60C).
For more information refer to ANSI/ISA 12.27.01-2003, the 2006 NEC (clauses 501.15
(F) (3) (CI Div 1/2), and 505.16 (E) (3) (CI Zone 1/2)) or the 2006 CEC Part 1 Rules
18-092 (CI Zone 0), 18-108 (CI Zone 1), 18-158 (CI Zone 2), J18-108 (CI Div1), and
J18-158 (CI Div2).
SPECIAL WARNINGS AND INFORMATION
EQUIPMENT USED IN CLASS I, DIV 2 AND ZONE 2 HAZARDOUS LOCATIONS
Refer to Chapter 2 – Specifications for details about the suitability of this equipment in hazard-
ous areas.
Explosion Hazard – Substitution of components may impair suitability for
hazardous locations.
Risque d’explosion – La substitution de composants peut rendre ce materiel
inacceptable pour les emplacements sont designés dangereux.
Explosion Hazard – Do not disconnect equipment unless power has been
switched off or the area is known to be non-hazardous.
Risque d’explosion – Avant de déconnecter l’équipement, coupez le courant où
vousassurez que l’emplacement est designé non dangereux.
All input and output wiring must be in accordance with Class I, Division 2 and Zone 2 wiring methods (NEC
Sec 501.4(b) or CEC 18-152) and in accordance with the authority having jurisdiction.
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...