6-6 | Model 5100 HD Analyzer
29
116
r
int16
MANCMD
82
1 Parameter used to request a state
machine change. 1=Not ready,
2=Measure, 3=Zero, 4=Span, 5=Zero
Hold, 6=Span Hold, 7=Idle
117
rw
char[20]
CONCUNIT
83
10 Concentration Unit of Measure for
analyte 1
118
rw
int16
FLOATFRMT
93
1 Float Format index. 0=comma as
decimal, 1=period as decimal
119
rw
uint16
IDATEFRMT
94
1 International date format index.
0=MM\DD\
YYYY, 1=MM-
DD
-YYYY,
2=DD\MM\
YYYY, 3=DD-
MM
-YYYY
120
rw
char[10]
PRESUNIT
95
5 Pressure Unit of Measure, case
sensitive. Valid strings include psi,
psia, inH20, inHg, bar, cmH20,
kg/cm2, kPa, mbar, micron, mmHg,
oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5 Temp Unit of Measure, case
sensistive. Valid strings include
degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2 Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-
lock mode, 1=Sample,
2=Calib
124
rwf
int16
COMPPT
108
1 Pressure and temp compensation.
0=none, 1=pressure, 2=temperature,
3=both
125
r
uint
SYSALARM1
109
2 Bit
-
mapped system alarms
126
r
uint
SYSALARM2
111
2 Bit
-
mapped system alarms
127
r
uint
USERALARM
113
2 Bit
-
mapped system alarms
128
r
uint
CONCALARM
115
2 Bit mapped concentration alarms
129
r
int16
SYSSTATUS
117
1
System Status. 0=Everything ok, Not
0=something is wrong
130
r
int16
SYSSTATE
118
1
System State. 100=Not Ready,
201=Measure Run, 502=Idle,
600=diag
131
r
uint
TOTUPTIME
119
2 Total uptime in seconds
132
r
uint
UPTIME
121
2 Uptime in seconds
133
r
float
CELLTEMPD
123
2 Cell temp for display
134
r
float
ANAOUT1
125
2
Current Output 1
135
rw
float
ANAOUT1O
127
2
Current Output 1 Offset
136
rw
uint16
ANAOUT1P
129
1
Current Output 1 Parameter
137
rw
float
ANAOUT1S
130
2
Current Output 1 Span
138
rw
int16
ANAOUT1T
132
1
Current Output 1 Output Type. 0=4-
20 mode, 1=0
-
20 mode.
If overange enable (ANAOUT1OR =
1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or
low values can also be set by fault
value, See Fault option (ANAOUT1F)
139
rw
uint16
ANAOUT1OR
133
1
Current Output 1 Over range Enable.
0=disabled, 1=enabled
140
rw
uint16
ANAOUT1F
134
1
Current Output 1 Fault Option. 0=No
fault, 1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT1T = 0):
Low end of range becomes 3 mA
Output ratio bec
omes 0.
30
141
r
float
ANAOUT2
135
2
Current Output 2 -
142
rw
float
ANAOUT2O
137
2
Current Output 2 Offset -
143
rw
uint16
ANAOUT2P
139
1
Current Output 2 Parameter-
144
rw
float
ANAOUT2S
140
2
Current Output 2 Span -
145
rw
int16
ANAOUT2T
142
1
Current Output 2 Output Type. 0=4-
20 mode, 1=0
-
20 mode.
If overange enable (ANAOUT2OR =
1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or
low values can also be set by fault
value, See Fault option (ANAOUT2F)
-
146
rw
uint16
ANAOUT2OR
143
1
Current Output 2
Over range Enable.
0=disabled, 1=enabled
147
rw
uint16
ANAOUT2F
144
1
Current Output 2 Fault Option. 0=No
fault, 1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT2T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
148
r
uint16
RELAY1
145
1
Relay 1. 0=OPEN, 1=CLOSED
149
rw
uint16
RELAY1D
146
1
Relay 1 Delay. Holds the relay output
for this set amount of time.
150
rw
uint16
RELAY1F
147
1
Relay 1 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
151
rw
uint16
RELAY1S
148
1
Relay 1 Source. The source
parameter number used to evaluated
the relay function.
152
rw
float
RELAY1V
149
2
Relay 1 Value. The value used
to
evaluate the relay function.
153
r
uint16
RELAY2
151
1
Relay 2. 0=OPEN, 1=CLOSED
154
rw
uint16
RELAY2D
152
1
Relay 2 Delay. Holds the relay output
for this set amount of time.
155
rw
uint16
RELAY2F
153
1
Relay 2 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
156
rw
uint16
RELAY2S
154
1
Relay 2 Source. The source
parameter number used to evaluated
the relay function.
157
rw
float
RELAY2V
155
2
Relay 2 Value. The value used to
evaluate the relay function.
158
r
uint16
RELAY3
157
1
Relay 3. 0=OPEN, 1=CLOSED
159
rw
uint16
RELAY3D
158
1
Relay 3 Delay. Holds the relay output
for this set amount of time.
160
rw
uint16
RELAY3F
159
1
Relay 3 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
161
rw
uint16
RELAY3S
160
1
Relay 3 Source. The source
parameter number used to evaluated
the relay function.
162
rw
float
RELAY3V
161
2
Relay 3 Value. The value used to
evaluate the relay function.
163
r
uint16
RELAY4
163
1
Relay 4. 0=OPEN, 1=CLOSED
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...