6-18 | Model 5100 HD Analyzer
42
154
rw
uint16
RELAY2D
152
1
Relay 2 Delay. Holds the relay output for this
set amount of time.
155
rw
uint16
RELAY2F
153
1
Relay 2 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
156
rw
uint16
RELAY2S
154
1
Relay 2 Source. The source parameter
number used to evaluated the relay function.
157
rw
float
RELAY2V
155
2
Relay 2 Value. The value used to evaluate
the relay function.
158
r
uint16
RELAY3
157
1
Relay 3. 0=OPEN, 1=CLOSED
159
rw
uint16
RELAY3D
158
1
Relay 3 Delay. Holds the relay output for this
set amount of time.
160
rw
uint16
RELAY3F
159
1
Relay 3 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
161
rw
uint16
RELAY3S
160
1
Relay 3 Source. The source parameter
number used to evaluated the relay function.
162
rw
float
RELAY3V
161
2
Relay 3 Value. The value used to evaluate
the relay function.
163
r
uint16
RELAY4
163
1
Relay 4. 0=OPEN, 1=CLOSED
164
rw
uint16
RELAY4D
164
1
Relay 4 Delay. Holds the relay output for this
set amount of time.
165
rw
uint16
RELAY4F
165
1
Relay 4 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
166
rw
uint16
RELAY4S
166
1
Relay 4 Source. The source parameter
number used to evaluated the relay function.
167
rw
float
RELAY4V
167
2
Relay 4 Value. The value used to evaluate
the relay function.
168
rwf
boolean
RELAYENA
169
1 Master enable for all relays. 0=Disabled,
1=Enabled
169
rwf
char[20]
ALYTE1
170
10
Analyte 1 Name
170
r
float
ALYTE1CON
180
2
Analyte 1 Concentration
179
rwf
char[20]
ALYTE2
198
10
Analyte 2 Name
180
r
float
ALYTE2CON
208
2
Analyte 2 Concentration
181
r
float
TECTEMP
210
2
TEC Temperature Value
192
r
float
CELLTEMP
231
2 Cell Temperature value
200
rw
char[10]
CONCUNIT2
247
5 Concentration Unit of Measure for analyte 2
201
rw
uint
ALYTE1RES
252
2 Decimal places to use for displaying analyte 1
202
rw
uint
ALYTE2RES
254
2 Decimal places to use for displaying analyte 2
203
rw
boolean
RELAY1NC
256
1 Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
204
rw
boolean
RELAY2NC
257
1 Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
205
rw
boolean
RELAY3NC
258
1 Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
206
rw
boolean
RELAY4NC
259
1 Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
209
r
uint
SYSWARN1
262
2 Temporary cache to miror warnings in
SYSALARM1
210
r
uint
SYSWARN2
264
2 Temporary cache to miror warnings in
SYSALARM2
211
r
uint
SYSSTWARN
266
2 system statuc warning
215
rw
float
REFCONC_2
273
2
Concentration of Reference call
219
rw
int16
ANAOUT1H
280
1
Analog output 1, 0 -
follow source during cal,
1
-
hold output during cal
220
rw
int16
ANAOUT2H
281
1
Analog output 2, 0 -
follow source during cal,
1
-
hold output during cal L82
43
227
rw
float
CONCMIN_2
293
2 Minimum Concentration
228
rw
float
CONCMAX_2
295
2 Maximum Concentration
234
rwf
float
REFREFCONC
306
2
Reference
concentration of the reference cell
at
235
rw
boolean
PPSRC
308
1
if FALSE, use static pressure
236
rw
float
PPMAX
309
2 Max Pressure allowed in PPMV to Dew Point
237
rw
float
PPMIN
311
2 Min Pressure allowed in PPMV to Dew Point
239
rwf
float
REFCONCTOL
314
2
Reference concentration alarm band (%)
240
rwf
float
REFCONCTOL_2
316
2
Reference concentration alarm band 2 (%)
241
rwf
float
BGNDFACTOR
318
2 Background compensation factor
242
rwf
float
BGNDNORMTEMP
320
2 Background compensation temperature
243
r
float
CELLPRESD
322
2
User Analog Input 1 output (engr units)
244
r
float
PRESST
324
2 Transmiter Pressue
248
rw
float
FIXEDPRES
332
2
User Analog Input Fixed Substitute Value
249
rwf
float
CELLTEMPIIR
334
2
Cell RTD1 IIR Filter
Weight
250
rwf
float
REFTEMP
336
2
RTD2 (reference cell temperature)
252
r
float
ALYTE1CON_UC
340
2 Unclamped analyte 1 concentration value
253
r
float
ALYTE2CON_UC
342
2 Unclamped analyte 2 concentration value
254
r
float
CELLPRES
344
2
PRESX1 Analog
Input Filtered value
273
rw
float
CONCMIN
379
2 Minimum Concentration
274
rw
float
CONCMAX
381
2 Maximum Concentration
344
rwf
float
REFCONC
484
2
Concentration of Reference call
361
r
float
ALYTE1CONS
517
2
Analyte 1 Concentration in other unit
362
rw
char[20]
CONCUNITS
519
10 2nd Concentration Unit of Measure for
analyte 1
363
r
uint
DATAVALID_2
529
2 Data Valid. 0=Not Valid, 1=Valid
364
r
int16
DMODE_2
531
1
0=Line-
lock mode, 1=Sample, 2=Calib
369
rw
float
SEALPRESMAX
539
2 Process seal max
370
rw
float
SEALPRESMIN
541
2 Process seal min
374
r
float
TECTEMP_2
549
2
TEC Temperature Value
395
rwf
float
TEMPREF_2
585
2
Reference temperature for this diode
430
rw
float
CONCSPAN
651
2 concentration span
431
rw
float
CONCZERO
653
2 concentration offset
432
rw
float
CONCSPAN_2
655
2 concentration span 2
433
rw
float
CONCZERO_2
657
2 concentration offset 2
452
rw
int16
AUTOCAL
687
1
Auto calibration. 0=disabled, 1=enabled.
453
rw
int16
CALMODE
688
1 0=Zero+Span, 1=Zero, 2=Span
454
rw
int16
CALDAY
689
1 The last day that the calibration was triggered
by the scheduler
455
rw
int16
CALPERIOD
690
1 Cal period in second
456
r
uint
CALSTART
691
2 1=perform autocal immediately
457
r
float
ALYTE3CON
693
2
Analyte 3 (merged) concentration
458
r
float
ALYTE3CON_UC
695
2 Unclamped analyte 1 concentration value
459
rw
int16
AUTOADJUST
697
1 0=do not adjust auto cal; 1=adjust auto cal
460
rwf
int16
USESTDDEV
698
1
User standard deviation in WIR filter
461
r
int16
SYSSTATE_2
699
1 System state 2.
100=Not Ready,
201=Measure Run, 502=Idle, 600=diag
462
rwf
float
REFTEMP2
700
2
Reference temperature 2
473
rwf
int16
ATEX
718
1
Boolean, 1=ATEX, 0=normal
474
r
float
SEALPRES
719
2 Process seal presure
494
rwf
float
REFCONCOMP
753
2
Reference
compensation
495
rwf
float
CELLPRESMAX
755
2 Cell pressure hi limit
496
rwf
float
CELLPRESMIN
757
2 Cell pressure lo limit
497
r
float
ALYTE3MCON
759
2
Analyte 3 (merged) concentration
498
rwf
float
CRSOVRCONC
761
2 Cross
-over point for Alyte 3
concentration
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...