System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) contains none-zero value,
the bit map of System Alarms and System Warning (parameter 125, 126, 127, 128; 32-bit integer,
modbus register 109, 110, 111, 112, 113) contains system alarms as defined:
Modbus
register
Bit Display message
Alarm
type
Related
Parameter ID
Description
109
1
“TEC Hi Limit Alrm”
Alarm
181, 183, 188
Temperature of the TEC#1 (181) is
outside of a band set by parameter 183
around the setpoint (188)
109
2
“TEC Lo Limit Alrm”
Alarm
109
3
“TEC2 Hi Limit Alrm”
Alarm
374, 376, 381
Temperature of the TEC#2 (374) is
outside of a band set by parameter 376
around the setpoint (381)
109
4
“TEC2 Lo Limit Alrm”
Alarm
109
5
“Cell Hi Limit Alrm”
Alarm
192, 193, 194
Cell temperature (192) is outside of a
band set by parameter 194 around the
setpoint (193)
109
6
“Cell Lo Limit Alrm”
Alarm
109
7
“Pres Hi Limit Alrm”
Alarm
244, 236, 237
Pressure transducer reading (244) is
outside of a band set by min/max
pressure (236, 237)
109
8
“Pres Lo Limit Alrm”
Alarm
109
9
“Seal Hi Limit Alrm”
Alarm
366, 369, 370
For ATEX ONLY: SEAL pressure (366)
is outside of a band set by min/max
pressure (369, 370)
109
10
“Seal Lo Limit Alrm”
Alarm
109
11 "File error"
Alarm
File system error, reported in system
alarm (125)
109
12 "Internal error"
Alarm
Internal harware error
111
13
“TEC temp drift”
Warning
181, 197, 198,
276
TEC temperature
(181)
drifts out of
factory setting window
(197, 198)
defined by the factory default (276)
111
14
“TEC2 temp drift”
Warning
381, 213, 214,
395
TEC temperature 2
(381)
drifts out of
factory setting window
(213, 214)
defined by the factory default (395)
111
15
"Conc. Out of range"
Warning
Concentration reading from cell #1 is out
of calibration range
111
16
"Conc. Out of range 2"
Warning
Concentration reading from cell #2 is
out of calibration range
111
17
"Lowlight in Sample"
Warning
Laser intensity in sample cell is low
111
18 "Battery failed"
Warning
Internal battery failure
111
19
"Lowlight in Ref"
Warning
Laser
intensity in reference cell is low
111
20
"Ref conc alarm"
Warning
Concentration in reference cell is low
111
21 "Conc. Band 1"
Warning
Concentration alarm from band 1
(channel 1)
111
22 "Conc. Band 2"
Warning
Concentration alarm from band 2
(channel 1)
111
23 "Conc. Band 3"
Warning
Concentration alarm from band 3
(channel 1)
111
24 "Conc. Band 4"
Warning
Concentration alarm from band 4
(channel 1)
111
25 "Conc. Band 1"
Warning
Concentration alarm from band 1
(channel 2)
111
26 "Conc. Band 2"
Warning
Concentration alarm from band 2
(channel 2)
111
27 "Conc. Band 3"
Warning
Concentration alarm from band 3
(channel 2)
111
28 "Conc. Band 4"
Warning
Concentration alarm from band 4
(channel 2)
111
29 "Data Invalid"
Warning
122, 363
Concentration reading is invalid.
Possible causes include: system in line
-
lock, system in calibration, etc...
System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) con
-
tains none-zero value, the bit map of System Alarms and System Warning (pa
-
rameter 125, 126, 127, 128; 32-bit integer, modbus register 109, 110, 111, 112,
113) contains system alarms as defined:
40
111
30
“Cell pres Hi”
Warning
254, 495, 496
Cell pressure (254) is outside of a band
set by min/max pressure (495, 496)
111
31
“Cell pres Low”
Warning
111
32 "Span error"
Warning
Error in span calibration
112
1 "Zero error"
Warning
Error in zero calibration
113
2 "User error"
Warning
User error detected from the user analog
input
Common Modbus Registers
ID
Access Data
Type
mnemonic
MB
Starting
Regs
MB Reg
Size
Description
6
rs
char[20]
SERIAL
0
10 Serial Number
7
rs
char[6]
MAC
10
3
MAC address
8
rws
uint
IP
13
2
IP Address
9
rws
uint
SUBNET
15
2 IP Subnet Mask
10
rws
uint
GW
17
2 Gateway
11
rws
uint
DNS
19
2 DNS
12
rws
uint16
DHCPE
21
1 DHCP enable. 1=disabled, 2=enabled
13
rws
uint16
TCPE
22
1
TCP Enable. 0=disabled, 1=enabled
14
rws
uint
SVRIP
23
2 Server IP
100
rwf
int16
CFGFILEVER
25
1 Version of the params.xml file generated by
this
101
r
char[20]
VERSION
26
10 Software Version
102
r
char[20]
NAME
36
10 Model Name
103
rw
uint
IU1BAUD
46
2
Internal UART1baud rate (MODBUS)
104
rwf
uint
IU2BAUD
48
2
Internal UART2 baud rate (DISPLAY)
105
rwf
uint
EU1BAUD
50
2
External UART1 baud rate (RS232)
106
rwf
uint
EU2BAUD
52
2
External UART2 baud rate
107
r
uint16
FWVERHRB
54
1 Interface Board FW version
108
r
uint16
CARDID
55
1 0x20: HC121, 0x40: HC122
109
rw
int16
MBPARITY
56
1 Parity. 0=none, 1=even, 2=odd
110
rwf
int16
MBSTOPBIT
57
1 Stop bits. 1=1 stop bits, 2=2 stop bits
111
rw
uint16
MBUSADD
58
1
Modbus Address
112
rwf
uint16
MBUSPORT
59
1
Modbus Port. 0=Disable, 1=Enable
113
rwf
uint
MBUSTO
60
2
Modbus Timeout (ms)
114
r
char[20]
SYSTIME
62
10 System Time
115
r
char[20]
SYSDATE
72
10 System Date
117
rw
char[20]
CONCUNIT
83
10 Concentration Unit of Measure for analyte 1
120
rw
char[10]
PRESUNIT
95
5 Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
mmHg, oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5 Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2 Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-
lock mode, 1=Sample, 2=Calib
125
r
uint
SYSALARM1
109
2 Bit
-
mapped system alarms
126
r
uint
SYSALARM2
111
2 Bit
-
mapped system alarms
127
r
uint
USERALARM
113
2 Bit
-
mapped system alarms
128
r
uint
CONCALARM
115
2 Bit mapped concentration alarms
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...