3-2 | Model 5100 HD Analyzer
Analyzer Requirements
Location
The Model 5100 HD Analyzer is wall-mounted; ensure that there is adequate clearance
when mounting and select a readily accessible position for the analyzer to allow for
routine maintenance. The location should be free from excessive vibration and the
ambient temperature should be -20 °C to 50 °C (-4 °F to 122 °F) without external
heating or cooling. Shade must be provided for the analyzer.
Do not mount in direct
sunlight. Shade from luminaires.
• Allow a minimum of 8 inches (20.3 cm) on both sides and the bottom of the
unit for space for system output cables and the AC mains supply conduit.
• Allow a minimum of 4 inches (10.2 cm) on the top of the unit to allow
access to mounting hardware should the unit need to be removed from the
wall.
• Provide adequate clearance at the bottom and sides of the unit for routing
of gas supply line and vent lines.
• The access door(s) to the enclosures swing out when open. Be sure to allow
space for door swing.
Mounting Requirements
Avoid mounting the analyzer in extremely dirty environments or where
it will be subject to large amounts of electromagnetic interference.
Mounting Hardware: (6) 5/16” x L (M8 x 25 mm L) hex bolts
NOTE
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...