Maintenance
|
6-17
41
129
r
int16
SYSSTATUS
117
1 System Status.
0=Everything ok, Not
0=something is wrong
130
r
int16
SYSSTATE
118
1
System State. 100=Not Ready, 201=Measure
Run, 502=Idle, 600=diag
133
r
float
CELLTEMPD
123
2 Cell temp for display
134
r
float
ANAOUT1
125
2
Current Output 1
135
rw
float
ANAOUT1O
127
2
Current Output 1 Offset
136
rw
uint16
ANAOUT1P
129
1
Current Output 1 Parameter
137
rw
float
ANAOUT1S
130
2
Current Output 1 Span
138
rw
int16
ANAOUT1T
132
1
Current Output 1 Output Type. 0=4-
20 mode,
1=0
-
20 mode.
If overange enable (ANAOUT1OR = 1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT1F)
139
rw
uint16
ANAOUT1OR
133
1
Current Output 1 Over range Enable.
0=disabled, 1=enabled
140
rw
uint16
ANAOUT1F
134
1
Current Output 1 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT1T = 0):
Low end of range becomes 3 mA
Output ratio bec
omes 0.
141
r
float
ANAOUT2
135
2
Current Output 2 -
142
rw
float
ANAOUT2O
137
2
Current Output 2 Offset -
143
rw
uint16
ANAOUT2P
139
1
Current Output 2 Parameter-
144
rw
float
ANAOUT2S
140
2
Current Output 2 Span -
145
rw
int16
ANAOUT2T
142
1
Current Output 2 Output Type. 0=4-
20 mode,
1=0
-
20 mode.
If overange enable (ANAOUT2OR = 1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT2F) -
146
rw
uint16
ANAOUT2OR
143
1
Current Output 2 Over range Enable.
0=disabled, 1=enabled
147
rw
uint16
ANAOUT2F
144
1
Current Output 2 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT2T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
148
r
uint16
RELAY1
145
1
Relay 1. 0=OPEN, 1=CLOSED
149
rw
uint16
RELAY1D
146
1
Relay 1 Delay. Holds the relay output for this
set amount of time.
150
rw
uint16
RELAY1F
147
1
Relay 1 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
151
rw
uint16
RELAY1S
148
1
Relay 1 Source. The source parameter
number used to evaluated the relay function.
152
rw
float
RELAY1V
149
2
Relay 1 Value. The value used to
evaluate
the relay function.
153
r
uint16
RELAY2
151
1
Relay 2. 0=OPEN, 1=CLOSED
42
154
rw
uint16
RELAY2D
152
1
Relay 2 Delay. Holds the relay output for this
set amount of time.
155
rw
uint16
RELAY2F
153
1
Relay 2 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
156
rw
uint16
RELAY2S
154
1
Relay 2 Source. The source parameter
number used to evaluated the relay function.
157
rw
float
RELAY2V
155
2
Relay 2 Value. The value used to evaluate
the relay function.
158
r
uint16
RELAY3
157
1
Relay 3. 0=OPEN, 1=CLOSED
159
rw
uint16
RELAY3D
158
1
Relay 3 Delay. Holds the relay output for this
set amount of time.
160
rw
uint16
RELAY3F
159
1
Relay 3 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
161
rw
uint16
RELAY3S
160
1
Relay 3 Source. The source parameter
number used to evaluated the relay function.
162
rw
float
RELAY3V
161
2
Relay 3 Value. The value used to evaluate
the relay function.
163
r
uint16
RELAY4
163
1
Relay 4. 0=OPEN, 1=CLOSED
164
rw
uint16
RELAY4D
164
1
Relay 4 Delay. Holds the relay output for this
set amount of time.
165
rw
uint16
RELAY4F
165
1
Relay 4 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
166
rw
uint16
RELAY4S
166
1
Relay 4 Source. The source parameter
number used to evaluated the relay function.
167
rw
float
RELAY4V
167
2
Relay 4 Value. The value used to evaluate
the relay function.
168
rwf
boolean
RELAYENA
169
1 Master enable for all relays. 0=Disabled,
1=Enabled
169
rwf
char[20]
ALYTE1
170
10
Analyte 1 Name
170
r
float
ALYTE1CON
180
2
Analyte 1 Concentration
179
rwf
char[20]
ALYTE2
198
10
Analyte 2 Name
180
r
float
ALYTE2CON
208
2
Analyte 2 Concentration
181
r
float
TECTEMP
210
2
TEC Temperature Value
192
r
float
CELLTEMP
231
2 Cell Temperature value
200
rw
char[10]
CONCUNIT2
247
5 Concentration Unit of Measure for analyte 2
201
rw
uint
ALYTE1RES
252
2 Decimal places to use for displaying analyte 1
202
rw
uint
ALYTE2RES
254
2 Decimal places to use for displaying analyte 2
203
rw
boolean
RELAY1NC
256
1 Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
204
rw
boolean
RELAY2NC
257
1 Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
205
rw
boolean
RELAY3NC
258
1 Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
206
rw
boolean
RELAY4NC
259
1 Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
209
r
uint
SYSWARN1
262
2 Temporary cache to miror warnings in
SYSALARM1
210
r
uint
SYSWARN2
264
2 Temporary cache to miror warnings in
SYSALARM2
211
r
uint
SYSSTWARN
266
2 system statuc warning
215
rw
float
REFCONC_2
273
2
Concentration of Reference call
219
rw
int16
ANAOUT1H
280
1
Analog output 1, 0 -
follow source during cal,
1
-
hold output during cal
220
rw
int16
ANAOUT2H
281
1
Analog output 2, 0 -
follow source during cal,
1
-
hold output during cal L82
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...