Maintenance
|
6-9
32
202
rw
uint
ALYTE2RES
254
2 Decimal places to use for displaying
analyte 2
203
rw
boolean
RELAY1NC
256
1 Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
204
rw
boolean
RELAY2NC
257
1 Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
205
rw
boolean
RELAY3NC
258
1 Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
206
rw
boolean
RELAY4NC
259
1 Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
207
rw
uint16
MENUTO
260
1 Menu timeout. The time that the menu
will navigate to the homescreen after
no activity. 0 is disabled
208
rw
uint16
PWDTO
261
1 Password timeout. The time that a
user logged into the front display
menu will be logged out after no
activity. 0 is disabled
209
r
uint
SYSWARN1
262
2 Temporary cache to miror warnings in
SYSALARM1
210
r
uint
SYSWARN2
264
2 Temporary cache to miror warnings in
SYSALARM2
211
r
uint
SYSSTWARN
266
2 system statuc warning
212
rw
uint16
PTSEED
268
1
The seed value for all IIR filters.
213
rwf
float
MAXTEMP_2
269
2
Max TEC temperature 2
214
rwf
float
MINTEMP_2
271
2
Min TEC temperature 2
215
rw
float
REFCONC_2
273
2
Concentration of Reference call
216
rwf
int16
STATICPT
275
1
Pressure (STCELLP) and temp
(STCELLT) static value usage.
0=neither, 1=use static pressure,
2=use static temperature, 3= use both
217
rwf
float
STCELLT
276
2 Static cell temperature. Use by
setting STATICPT to 2 or 3
218
rwf
float
STCELLP
278
2 Static cell pressure. Use by setting
STATICPT to 1 or 3
219
rw
int16
ANAOUT1H
280
1
Analog output 1, 0 -
follow source
during cal, 1
-
hold output during cal
220
rw
int16
ANAOUT2H
281
1
Analog output 2, 0 -
follow source
during cal, 1
-
hold output during cal
L82
221
rwf
float
LASTH_2
282
2 Threshold in 2f signal
222
rwf
float
LASREFPWR_2
284
2
Reference Laser Power (power at
calib)
223
rwf
float
LASTS_2
286
2
Laser Sample Period in seconds
224
rwf
float
LASPWR_2
288
2
Laser Power
225
rwf
float
CELLLENGTH
290
2 Cell length, concentrations are
multiplied by this parameter
226
r
int16
DIAGS
292
1 Used to change state to diagnostics
from the menu
227
rw
float
CONCMIN_2
293
2 Minimum Concentration
228
rw
float
CONCMAX_2
295
2 Maximum Concentration
229
rwf
int16
PASSCODE
297
1
PASSCODE to change system
params
230
rwf
float
LASTH
298
2 Threshold in 2f signal
231
rwf
float
LASREFPWR
300
2
Reference Laser Power (power at
calib)
232
rwf
float
LASTS
302
2
Laser Sample Period in seconds
233
rwf
float
LASPWR
304
2
Laser Power
33
234
rwf
float
REFREFCONC
306
2
Reference concentration of the
reference cell at
235
rw
boolean
PPSRC
308
1 if
FALSE, use static pressure
236
rw
float
PPMAX
309
2 Max Pressure allowed in PPMV to
Dew Point
237
rw
float
PPMIN
311
2 Min Pressure allowed in PPMV to
Dew Point
238
rwf
boolean
UAIFORM
313
1
Selects 0 offset formula if TRUE else
selects 4mA offset
formula for User
Analog Input
239
rwf
float
REFCONCTOL
314
2
Reference concentration alarm band
(%)
240
rwf
float
REFCONCTOL_2
316
2
Reference concentration alarm band
2 (%)
241
rwf
float
BGNDFACTOR
318
2 Background compensation factor
242
rwf
float
BGNDNORMTEMP
320
2 Background compensation
temperature
243
r
float
CELLPRESD
322
2
User Analog Input 1 output (engr
units)
244
r
float
PRESST
324
2 Transmiter Pressue
245
rw
float
PRESSTS
326
2
User Analog Input 1 Span
246
rw
float
PRESSTOS
328
2
User Analog Input 1 Offset
247
rwf
float
PRESSTIIR
330
2
Transmiter Pressue IIR Filter Weight
248
rw
float
FIXEDPRES
332
2
User Analog Input Fixed Substitute
Value
249
rwf
float
CELLTEMPIIR
334
2
Cell RTD1 IIR Filter Weight
250
rwf
float
REFTEMP
336
2
RTD2 (reference cell temperature)
251
rwf
float
REFTEMPIIR
338
2
RTD2 Analog Input Filter Weight
252
r
float
ALYTE1CON_UC
340
2 Unclamped analyte 1 concentration
value
253
r
float
ALYTE2CON_UC
342
2 Unclamped analyte 2 concentration
value
254
r
float
CELLPRES
344
2
PRESX1 Analog Input Filtered value
255
rwf
float
CELLPRESS
346
2 Pressure Transducer 1 Scale
256
rwf
float
CELLPRESOS
348
2
Pressure Transducer 1 Offset
257
rwf
float
CELLPRESIIR
350
2
PRESX1 IIR filter weight
258
rwf
float
TECIIR
352
2
TEC IIR Filter Weight
259
r
float
UNUSEDHTR1
354
2
(Not used)
260
rwf
float
UNUSEDHTR1S
356
2
(Not used)
261
rwf
float
UNUSEDHTR1OS
358
2
(Not used)
262
r
float
UNUSEDHTR2
360
2
(Not used)
263
rwf
float
CONCOUTH1
362
2 Concentration out of range Hight 1
264
rwf
float
CONCOUTL1
364
2
Concentration out of range Low 1
265
r
float
UNUSEDHTR3
366
2
(Not used)
266
rwf
float
CONCOUTH2
368
2 Concentration out of range Hight 1
267
rwf
float
CONCOUTL2
370
2 Concentration out of range
Low 1
268
rwf
int16
WLTOL1
372
1 Wavelength Tolerance 1
269
rwf
int16
WLTOL2
373
1 Wavelength Tolerance 2
270
rwf
float
IOFFSET
374
2 Injection current offset
271
rwf
float
LSTWLRES
376
2 Temperature Wavelength resolution
272
rwf
int16
MEDSIZE
378
1 Median filter size
273
rw
float
CONCMIN
379
2 Minimum Concentration
274
rw
float
CONCMAX
381
2 Maximum Concentration
275
r
int16
CTRINDX
383
1 Spectrum center index
276
rwf
float
TEMPREF
384
2
Reference temperature for this diode
277
rwf
int16
CONCSTART
386
1 Starting point of concentration window
278
rwf
int16
CONCEND
387
1
End point of concentration window
279
rwf
int16
CONCPEAK
388
1 Peak index for linklock
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...