Maintenance
|
6-5
Parameter ID & Modbus Registers
All system parameters can be identified by a parameter ID. Telnet/Terminal inter
-
face use this ID to view, set, and trend the parameter’s value
Access: r: read, w: write, s: system, f: factory, h: hidden
Parameter 112: 0 = Modbus disabled, 1=Modbus RTU enabled, 4=Modbus TCP/
IP enabled.
28
Parameter ID & Modbus Registers
All system parameters can be identified by a parameter ID. Telnet/Terminal interface use this ID to
view, set, and trend the parameter’s value
Access: r: read, w: write, s: system, f: factory, h: hidden
Parameter 112: 0 = Modbus disabled, 1=Modbus RTU enabled, 4=Modbus TCP/IP enabled
ID
Access
Data
Type
mnemonic
MB Start
Regs
MB Reg
Size
Description
6
rs
char[20]
SERIAL
0
10 Serial Number
7
rs
char[6]
MAC
10
3
MAC address
8
rws
uint
IP
13
2
IP Address
9
rws
uint
SUBNET
15
2 IP Subnet Mask
10
rws
uint
GW
17
2 Gateway
11
rws
uint
DNS
19
2 DNS
12
rws
uint16
DHCPE
21
1 DHCP enable. 1=disabled, 2=enabled
13
rws
uint16
TCPE
22
1
TCP Enable. 0=disabled, 1=enabled
14
rws
uint
SVRIP
23
2 Server IP
100
rwf
int16
CFGFILEVER
25
1 Version of the params.xml file
generated by this
101
r
char[20]
VERSION
26
10 Software Version
102
r
char[20]
NAME
36
10 Model Name
103
rw
uint
IU1BAUD
46
2
Internal UART1baud rate (MODBUS)
104
rwf
uint
IU2BAUD
48
2
Internal UART2 baud rate (DISPLAY)
105
rwf
uint
EU1BAUD
50
2
External UART1 baud rate (RS232)
106
rwf
uint
EU2BAUD
52
2
External UART2 baud rate
107
r
uint16
FWVERHRB
54
1 Interface Board FW version
108
r
uint16
CARDID
55
1 0x20: HC121, 0x40: HC122
109
rw
int16
MBPARITY
56
1 Parity. 0=none, 1=even, 2=odd
110
rwf
int16
MBSTOPBIT
57
1 Stop bits. 1=1 stop bits, 2=2 stop bits
111
rw
uint16
MBUSADD
58
1
Modbus Address
112
rwf
uint16
MBUSPORT
59
1
Modbus Port. 0=Disable, 1=RTU,
4=TCP/IP
113
rwf
uint
MBUSTO
60
2
Modbus Timeout (ms)
114
r
char[20]
SYSTIME
62
10 System Time
115
r
char[20]
SYSDATE
72
10 System Date
Chapter
10
29
116
r
int16
MANCMD
82
1 Parameter used to request a state
machine change. 1=Not ready,
2=Measure, 3=Zero, 4=Span, 5=Zero
Hold, 6=Span Hold, 7=Idle
117
rw
char[20]
CONCUNIT
83
10 Concentration Unit of Measure for
analyte 1
118
rw
int16
FLOATFRMT
93
1 Float Format index. 0=comma as
decimal, 1=period as decimal
119
rw
uint16
IDATEFRMT
94
1 International date format index.
0=MM\DD\
YYYY, 1=MM-
DD
-YYYY,
2=DD\MM\
YYYY, 3=DD-
MM
-YYYY
120
rw
char[10]
PRESUNIT
95
5 Pressure Unit of Measure, case
sensitive. Valid strings include psi,
psia, inH20, inHg, bar, cmH20,
kg/cm2, kPa, mbar, micron, mmHg,
oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5 Temp Unit of Measure, case
sensistive. Valid strings include
degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2 Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-
lock mode, 1=Sample,
2=Calib
124
rwf
int16
COMPPT
108
1 Pressure and temp compensation.
0=none, 1=pressure, 2=temperature,
3=both
125
r
uint
SYSALARM1
109
2 Bit
-
mapped system alarms
126
r
uint
SYSALARM2
111
2 Bit
-
mapped system alarms
127
r
uint
USERALARM
113
2 Bit
-
mapped system alarms
128
r
uint
CONCALARM
115
2 Bit mapped concentration alarms
129
r
int16
SYSSTATUS
117
1
System Status. 0=Everything ok, Not
0=something is wrong
130
r
int16
SYSSTATE
118
1
System State. 100=Not Ready,
201=Measure Run, 502=Idle,
600=diag
131
r
uint
TOTUPTIME
119
2 Total uptime in seconds
132
r
uint
UPTIME
121
2 Uptime in seconds
133
r
float
CELLTEMPD
123
2 Cell temp for display
134
r
float
ANAOUT1
125
2
Current Output 1
135
rw
float
ANAOUT1O
127
2
Current Output 1 Offset
136
rw
uint16
ANAOUT1P
129
1
Current Output 1 Parameter
137
rw
float
ANAOUT1S
130
2
Current Output 1 Span
138
rw
int16
ANAOUT1T
132
1
Current Output 1 Output Type. 0=4-
20 mode, 1=0
-
20 mode.
If overange enable (ANAOUT1OR =
1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or
low values can also be set by fault
value, See Fault option (ANAOUT1F)
139
rw
uint16
ANAOUT1OR
133
1
Current Output 1 Over range Enable.
0=disabled, 1=enabled
140
rw
uint16
ANAOUT1F
134
1
Current Output 1 Fault Option. 0=No
fault, 1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT1T = 0):
Low end of range becomes 3 mA
Output ratio bec
omes 0.
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...