4-22 | Model 5100 HD Analyzer
Data Type
Description
MB Starting
Regs
MB Reg Size
int16
0=Line-lock mode, 1=Sample, 2=Calib
107
1
int16
Pressure and temp compensation. 0=none,
1=pressure, 2=temperature, 3=both
108
1
uint
Bit-mapped system alarms
109
2
uint
Bit-mapped system alarms
111
2
uint
Bit-mapped system alarms
113
2
uint
Bit mapped concentration alarms
115
2
int16
System Status. 0=Everything ok, Not 0=some-
thing is wrong
117
1
int16
System State. 100=Not Ready, 201=Measure
Run, 502=Idle, 600=diag
118
1
uint
Total uptime in seconds
119
2
uint
Uptime in seconds
121
2
float
Cell temp for display
123
2
float
Current Output 1
125
2
float
Current Output 1 Offset
127
2
uint16
Current Output 1 Parameter
129
1
float
Current Output 1 Span
130
2
int16
Current Output 1 Output Type. 0=4-20 mode,
1=0-20 mode.
If overange enable (ANAOUT1OR = 1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low val-
ues can also be set by fault value, See Fault
option (ANAOUT1F)
132
1
uint16
Current Output 1 Over range Enable. 0=dis-
abled, 1=enabled
133
1
uint16
Current Output 1 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-20 mode (ANAOUT1T
= 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
134
1
float
Current Output 2 -
135
2
float
Current Output 2 Offset -
137
2
uint16
Current Output 2 Parameter-
139
1
float
Current Output 2 Span -
140
2
int16
Current Output 2 Output Type. 0=4-20 mode,
1=0-20 mode.
If overange enable (ANAOUT2OR = 1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low val-
ues can also be set by fault value, See Fault
option (ANAOUT2F) -
142
1
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...