Maintenance
|
6-19
44
499
r
int16
ALYTESEL
763
1
Analyte 3 (merged) concentration
500
rw
int16
CALTIME
764
1 Time until next cal cycle
501
rw
char[20]
CALSCHEDT
765
10 The auto calibration schedule.
-
1 is disabled.
The time should be a number like 14:30:45 for
2:30:45 PM.
502
r
char[20]
NEXTCAL
775
10 Next schedule for calibration
43
227
rw
float
CONCMIN_2
293
2 Minimum Concentration
228
rw
float
CONCMAX_2
295
2 Maximum Concentration
234
rwf
float
REFREFCONC
306
2
Reference
concentration of the reference cell
at
235
rw
boolean
PPSRC
308
1
if FALSE, use static pressure
236
rw
float
PPMAX
309
2 Max Pressure allowed in PPMV to Dew Point
237
rw
float
PPMIN
311
2 Min Pressure allowed in PPMV to Dew Point
239
rwf
float
REFCONCTOL
314
2
Reference concentration alarm band (%)
240
rwf
float
REFCONCTOL_2
316
2
Reference concentration alarm band 2 (%)
241
rwf
float
BGNDFACTOR
318
2 Background compensation factor
242
rwf
float
BGNDNORMTEMP
320
2 Background compensation temperature
243
r
float
CELLPRESD
322
2
User Analog Input 1 output (engr units)
244
r
float
PRESST
324
2 Transmiter Pressue
248
rw
float
FIXEDPRES
332
2
User Analog Input Fixed Substitute Value
249
rwf
float
CELLTEMPIIR
334
2
Cell RTD1 IIR Filter
Weight
250
rwf
float
REFTEMP
336
2
RTD2 (reference cell temperature)
252
r
float
ALYTE1CON_UC
340
2 Unclamped analyte 1 concentration value
253
r
float
ALYTE2CON_UC
342
2 Unclamped analyte 2 concentration value
254
r
float
CELLPRES
344
2
PRESX1 Analog
Input Filtered value
273
rw
float
CONCMIN
379
2 Minimum Concentration
274
rw
float
CONCMAX
381
2 Maximum Concentration
344
rwf
float
REFCONC
484
2
Concentration of Reference call
361
r
float
ALYTE1CONS
517
2
Analyte 1 Concentration in other unit
362
rw
char[20]
CONCUNITS
519
10 2nd Concentration Unit of Measure for
analyte 1
363
r
uint
DATAVALID_2
529
2 Data Valid. 0=Not Valid, 1=Valid
364
r
int16
DMODE_2
531
1
0=Line-
lock mode, 1=Sample, 2=Calib
369
rw
float
SEALPRESMAX
539
2 Process seal max
370
rw
float
SEALPRESMIN
541
2 Process seal min
374
r
float
TECTEMP_2
549
2
TEC Temperature Value
395
rwf
float
TEMPREF_2
585
2
Reference temperature for this diode
430
rw
float
CONCSPAN
651
2 concentration span
431
rw
float
CONCZERO
653
2 concentration offset
432
rw
float
CONCSPAN_2
655
2 concentration span 2
433
rw
float
CONCZERO_2
657
2 concentration offset 2
452
rw
int16
AUTOCAL
687
1
Auto calibration. 0=disabled, 1=enabled.
453
rw
int16
CALMODE
688
1 0=Zero+Span, 1=Zero, 2=Span
454
rw
int16
CALDAY
689
1 The last day that the calibration was triggered
by the scheduler
455
rw
int16
CALPERIOD
690
1 Cal period in second
456
r
uint
CALSTART
691
2 1=perform autocal immediately
457
r
float
ALYTE3CON
693
2
Analyte 3 (merged) concentration
458
r
float
ALYTE3CON_UC
695
2 Unclamped analyte 1 concentration value
459
rw
int16
AUTOADJUST
697
1 0=do not adjust auto cal; 1=adjust auto cal
460
rwf
int16
USESTDDEV
698
1
User standard deviation in WIR filter
461
r
int16
SYSSTATE_2
699
1 System state 2.
100=Not Ready,
201=Measure Run, 502=Idle, 600=diag
462
rwf
float
REFTEMP2
700
2
Reference temperature 2
473
rwf
int16
ATEX
718
1
Boolean, 1=ATEX, 0=normal
474
r
float
SEALPRES
719
2 Process seal presure
494
rwf
float
REFCONCOMP
753
2
Reference
compensation
495
rwf
float
CELLPRESMAX
755
2 Cell pressure hi limit
496
rwf
float
CELLPRESMIN
757
2 Cell pressure lo limit
497
r
float
ALYTE3MCON
759
2
Analyte 3 (merged) concentration
498
rwf
float
CRSOVRCONC
761
2 Cross
-over point for Alyte 3
concentration
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...