6-8 | Model 5100 HD Analyzer
31
164
rw
uint16
RELAY4D
164
1
Relay 4 Delay. Holds the relay output
for this set amount of time.
165
rw
uint16
RELAY4F
165
1
Relay 4 Function. 0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
166
rw
uint16
RELAY4S
166
1
Relay 4 Source. The source
parameter number used to evaluated
the relay function.
167
rw
float
RELAY4V
167
2
Relay 4 Value. The value used to
evaluate the relay function.
168
rwf
boolean
RELAYENA
169
1 Master enable for all relays.
0=Disabled, 1=Enabled
169
rwf
char[20]
ALYTE1
170
10
Analyte 1 Name
170
r
float
ALYTE1CON
180
2
Analyte 1 Concentration
171
rw
float
CONCLIMITH1
182
2
Analyte 1 band high limite band 1
172
rw
float
CONCLIMITL1
184
2
Analyte 1 band low limite band 1
173
rw
float
CONCLIMITH2
186
2
Analyte 1
band high limite band 2
174
rw
float
CONCLIMITL2
188
2
Analyte 1 band low limite band 2
175
rw
float
CONCLIMITH3
190
2
Analyte 1 band high limite band 3
176
rw
float
CONCLIMITL3
192
2
Analyte 1 band low limite band 3
177
rw
float
CONCLIMITH4
194
2
Analyte 1 band high limite band 4
178
rw
float
CONCLIMITL4
196
2
Analyte 1 band low limite band 4
179
rwf
char[20]
ALYTE2
198
10
Analyte 2 Name
180
r
float
ALYTE2CON
208
2
Analyte 2 Concentration
181
r
float
TECTEMP
210
2
TEC Temperature Value
182
rwf
int16
TECAT
212
1
Heater Actuator Type. 0=Reverse,
1=Forward.
183
rwf
float
TECDEVT
213
2 Heater Deviation Tolerance. Ignored if
0. Otherwise high or low limit alarm
are set when value is out bounds of 1
deviation
184
rwf
float
TECDT
215
2 Heater Deriviative Time
185
r
float
TECHDC
217
2 Heater Duty Cycle
186
rwf
float
TECIT
219
2 Heater Integral Time
187
rwf
float
TECPB
221
2 Heater Proportional Band
188
rwf
float
TECSP
223
2 Heater Set Point
189
rwf
float
TECST
225
2
TEC Sample Time
190
rwf
float
TECHYS
227
2
TEC hysterisi
191
r
float
ELECTEMP
229
2
Electronics Temperature
192
r
float
CELLTEMP
231
2
Cell RTD1 Filtered value
193
rwf
float
CELLSP
233
2 Cell Set Point
194
rwf
float
CELLDEVT
235
2 Cell Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are
set when value is out bounds of 1
deviation
195
rwf
float
NORMTEMP
237
2 Calibration temperature for
concentration compensation.
196
rwf
float
NORMPRES
239
2 Calibration pressure for concentration
compensation
197
rwf
float
MAXTEMP
241
2
Max TEC temperature
198
rwf
float
MINTEMP
243
2
Min TEC temperature
199
rw
uint
LOGSAVEINT
245
2 Interval in minutes to save the log
data
200
rw
char[10]
CONCUNIT2
247
5 Concentration Unit of Measure for
analyte 2
201
rw
uint
ALYTE1RES
252
2 Decimal places to use for displaying
analyte 1
32
202
rw
uint
ALYTE2RES
254
2 Decimal places to use for displaying
analyte 2
203
rw
boolean
RELAY1NC
256
1 Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
204
rw
boolean
RELAY2NC
257
1 Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
205
rw
boolean
RELAY3NC
258
1 Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
206
rw
boolean
RELAY4NC
259
1 Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
207
rw
uint16
MENUTO
260
1 Menu timeout. The time that the menu
will navigate to the homescreen after
no activity. 0 is disabled
208
rw
uint16
PWDTO
261
1 Password timeout. The time that a
user logged into the front display
menu will be logged out after no
activity. 0 is disabled
209
r
uint
SYSWARN1
262
2 Temporary cache to miror warnings in
SYSALARM1
210
r
uint
SYSWARN2
264
2 Temporary cache to miror warnings in
SYSALARM2
211
r
uint
SYSSTWARN
266
2 system statuc warning
212
rw
uint16
PTSEED
268
1
The seed value for all IIR filters.
213
rwf
float
MAXTEMP_2
269
2
Max TEC temperature 2
214
rwf
float
MINTEMP_2
271
2
Min TEC temperature 2
215
rw
float
REFCONC_2
273
2
Concentration of Reference call
216
rwf
int16
STATICPT
275
1
Pressure (STCELLP) and temp
(STCELLT) static value usage.
0=neither, 1=use static pressure,
2=use static temperature, 3= use both
217
rwf
float
STCELLT
276
2 Static cell temperature. Use by
setting STATICPT to 2 or 3
218
rwf
float
STCELLP
278
2 Static cell pressure. Use by setting
STATICPT to 1 or 3
219
rw
int16
ANAOUT1H
280
1
Analog output 1, 0 -
follow source
during cal, 1
-
hold output during cal
220
rw
int16
ANAOUT2H
281
1
Analog output 2, 0 -
follow source
during cal, 1
-
hold output during cal
L82
221
rwf
float
LASTH_2
282
2 Threshold in 2f signal
222
rwf
float
LASREFPWR_2
284
2
Reference Laser Power (power at
calib)
223
rwf
float
LASTS_2
286
2
Laser Sample Period in seconds
224
rwf
float
LASPWR_2
288
2
Laser Power
225
rwf
float
CELLLENGTH
290
2 Cell length, concentrations are
multiplied by this parameter
226
r
int16
DIAGS
292
1 Used to change state to diagnostics
from the menu
227
rw
float
CONCMIN_2
293
2 Minimum Concentration
228
rw
float
CONCMAX_2
295
2 Maximum Concentration
229
rwf
int16
PASSCODE
297
1
PASSCODE to change system
params
230
rwf
float
LASTH
298
2 Threshold in 2f signal
231
rwf
float
LASREFPWR
300
2
Reference Laser Power (power at
calib)
232
rwf
float
LASTS
302
2
Laser Sample Period in seconds
233
rwf
float
LASPWR
304
2
Laser Power
Содержание 5100 HD
Страница 23: ...Overview 1 11 Space Coupled Option Schematic...
Страница 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Страница 39: ...Installation Start Up 3 11 I O Signals...
Страница 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Страница 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Страница 120: ...6 20 Model 5100 HD Analyzer...
Страница 129: ...Appendix A A 9 check or bump is performed disable or a span zero adjustment is performed enable...